5. Password Menu 

[P18] Password 

This option prevents malicious and unskilled user from changing option 
values. Once this option is set, the password is required whenever entering 
into program mode. The initial password is “0.” 

 

 

1. If the password is not “0”, the “pass” shows on the 

display.  

 

2. After 1 second, it moves to setting mode 

automatically. The first digit flickers. Use “UP” and 

“DOWN” button to set the password. Use “MENU” 

button to move the digit position.   

 

3. After setting password, press “RUN” button, the 

password is saved and return to the program mode. 

 

4. When the password is incorrect, the error message 

shows on the display. After 1 second it returns to 

setting password mode. 

 

 

• 

User can’t configure the menus when user forgets the 
password. Please note password number and pay attention 
not to lose it. 

 

 

 

 

 

 

 

Summary of Contents for ULM-1000C

Page 1: ...IS Technologies Co Ltd PAGE 1 41 Please read this manual attentively before installation...

Page 2: ...t Inspection 6 5 Symbols 7 II PRODUCT 9 1 Principle of operation 10 2 Specification 11 3 Product Package 12 4 Dimension 12 III INSTALLATION 14 1 General Guide 14 2 Mounting 15 3 Installation Condition...

Page 3: ...AGE 3 41 VI PROGRAMING 27 1 Application Menu 27 2 Current Output Menu 29 3 Compensation Menu 31 4 Outputs Simulation Menu 34 5 Password Menu 35 VII MAINTENANCE 37 Warranty Period 37 Repair Service 37...

Page 4: ...package or it can be downloaded through our website www sondar com Please note that the contents of this manual are subject to change without prior notice if the product is modified upgraded or impro...

Page 5: ...IS Technologies Co Ltd PAGE 5 41 Safety Guide Instruction...

Page 6: ...s misuse or modification of the product without manufacture s permission Conduct periodic inspection of the product 3 Cautions This manual provides all information you need to operate ULM 1000C maint...

Page 7: ...ored faults or malfunctions could be result Warning If it is ignored injury to people and serious damage to the instrument could be result Electric Shock If it is ignored the product could be damaged...

Page 8: ...IS Technologies Co Ltd PAGE 8 41 Product Description ULM 1000C...

Page 9: ...ment range is 6 meters ULM 1000C menu makes the user easy and simple for calibration All functions are optimized which enables you to effectively monitor the liquid level and keep your facilities runn...

Page 10: ...e of the target and received back by the sensor The running time is converted into the distance and it shows as level or volume on the display D C T 2 D DISTANCE C SOUND VELOCITY T TIME OF FLIGHT Dist...

Page 11: ...1m min 100m min adjustable Output Analog Analog 4 20mA into max 60 0 Display 5 Digit LCD Material Polypropylene IP Rating IP67 Process Connection 2 PF Temperature 20 70 4 158 for Sensor 20 60 4 140 fo...

Page 12: ...IS Technologies Co Ltd PAGE 12 41 3 Product Package PG13 5 ULM 1000C Manual CD Cable Grand X1 Test Report 4 Dimension The enclosure material is polypropylene and the protection grade is IP67...

Page 13: ...IS Technologies Co Ltd PAGE 13 41 INSTALLATION...

Page 14: ...red target such as ladders limit switches heating spirals etc When mounting the sensor keep the distance to the vessel wall The bottom of the sensor should be perpendicular to the surface of water Do...

Page 15: ...ket or with screw in flange or in many other ways This manual explains two typical mounting types of ULM 1000C 2 1 Installation with mounting bracket If the equipment is installed or used in a manner...

Page 16: ...ew in a flange The nozzle length shouldn t be exceeded the maximum length above Shorter setting is better The nozzle should be free of obstructions such as weld seams Nozzle Diameter D Maximum Nozzle...

Page 17: ...or in or above a filling flow outlet Secure enough distance from the filling flow When mounting the sensor keep the distance to the vessel wall The bottom of the sensor should be perpendicular to the...

Page 18: ...IS Technologies Co Ltd PAGE 18 41 Wiring...

Page 19: ...If the equipment is installed or used in a manner not specified in this manual then the protection provided by the equipment may be impaired 2 Power The standard power type is DC power The power range...

Page 20: ...IS Technologies Co Ltd PAGE 20 41 Operation...

Page 21: ...GE 21 41 V Operation 1 Start up Display When ULM 1000C is powered on the screen shows as bellow pictures After showing Pic2 the actual level value is displayed on the display screen Pic1 Start up Disp...

Page 22: ...e in real time while it is operating Programing Mode is to calibrate the menus for individual applications The 4 digits is shown in the LCD in the Measuring Mode A Flashing 0 is shown when it is in fa...

Page 23: ...None All Off Display indicates flashing E Indicates that ULM 1000C has gone into Fail condition E0 means there is no reflected echo received E1 means transducer and 2 3 Measuring Mode This mode is use...

Page 24: ...to another menu 3 Buttons To access the Program Mode simply press the Mode button Confirmation that you have entered the Program Mode will be given by the Detect and Mode Level Distance Space LED s b...

Page 25: ...the S W version the menu number displays Use UP and DOWN button go to P04 Press menu button the first digit flickers With UP and DOWN button the number is changed After first digit setting press MENU...

Page 26: ...IS Technologies Co Ltd PAGE 26 41 Programing...

Page 27: ...n sets the variable of measurement when in measuring mode and can be set to one of the following Option Value Description 1 Level Display shows how full the vessel is with respect to the Empty 0 of Sp...

Page 28: ...he face of the transducer to the bottom distance in cm inch The tank has to be empty User can set factory set value with up and down keys The maximum set value is 600 for ULM 10C X06 1000 for ULM 10C...

Page 29: ...to 20mA This option sets the distance or level or space depending on the selected Operating Mode Option 01 at which the 4mA output will occur By default 4mA will represent Empty 0 of Span P07 20 mA S...

Page 30: ...indicates such a condition Option 1 3 8 mA Fault condition LOE indicated by 3 8mA 2 Hold The previous measured value outputs 3 22mA Fault condition LOE indicated by 22mA Percentage P09 mA Fail safe Ti...

Page 31: ...0 01m min Responds to changes to a max 0 01m min Slow Step 0 01 Minimum 0 01 Maximum 100 0 100 0 100m min Responds to changes to a max 100m min Fast P11 Detection Threshold Voltage This option determi...

Page 32: ...ocity This option allows for the velocity of sound to be changed according to the atmosphere the transducer is operating in By default the velocity is set for sound travelling in air at a temperature...

Page 33: ...ed with the actual level several times then Vapor Temperature Compensation adjusted accordingly to obtain an accurately displayed reading The pre set is 60cm o C P15 Detection Algorithm This option de...

Page 34: ...P17 Simulation Put the range at your disposal ULM 1000C sends out the output current for the range For example if you put 6 meter 0 600 the output current is sent out respond to 6 meter If the range...

Page 35: ...2 After 1 second it moves to setting mode automatically The first digit flickers Use UP and DOWN button to set the password Use MENU button to move the digit position 3 After setting password press R...

Page 36: ...IS Technologies Co Ltd PAGE 36 41 Maintenance...

Page 37: ...creen it shows what the problem is The error code information can be founded in this manual Even though conduct every process by the trouble shootings in the manual still the problem exists contract a...

Page 38: ...IS Technologies Co Ltd PAGE 38 41 Trouble shooting...

Page 39: ...se Barrier within the ultrasonic range 1 Set the velocity lower than current setting value or 2 Set the Detection Threshold Voltage lower than current voltage or 3 Check the sensor bottom surface if t...

Page 40: ...A Output 0 7 000m 6 000m P08 Error 1 3 8mA 2 Hold 3 22mA 22mA P09 mA Fail safe Time 20 900sec 300sec Compensation P10 Damping Rate 0 01 100 min 1m min P11 Threshold 3 Low 15 High 3 Low P12 Output Powe...

Page 41: ...nologies Co Ltd PAGE 41 41 IS Technologies Co Ltd 11 78 Songdo dong Yeonsu gu Incheon City 406 840 Republic of Korea Phone 82 32 850 2624 Fax 82 32 850 2612 Email sales sondar com IS Technologies Co L...

Reviews: