12
IRRA
flow
Laser Leveler User Manual
7000494 Rev C, USA
Cleaning and Disinfection
The Laser Leveler shall be cleaned and disinfected after each treatment.
If fluids spill onto the Laser Leveler during treatment, pause or stop treatment
and wipe off the spillage immediately.
The recommended method for cleaning is to wipe the parts with a soft cloth
dampened with soapy water until all foreign material has been removed.
The recommended method for surface disinfectant is to wipe the parts with
a soft cloth dampened with the following intermediate level disinfectant
agents or an intermediate level disinfectant wipe.
Only the following intermediate level disinfectant liquid may be used with a
minimum dwell time of one minute exposure:
70% Isopropyl alcohol or 70% ethyl alcohol solution
No liquid may be allowed to be directly poured or sprayed onto the Laser
Leveler during cleaning or disinfecting as this may damage the equipment.
No liquid may be allowed to dispel from the cloth during cleaning, as this
may damage the equipment.
Never use any type of tools or brush when cleaning, as this may damage the
equipment.
Device is supplied non sterile. No components, parts or accessories to the
IRRA
flow
Laser Leveler may be sterilized.
IRRAS AB does not make any claims for or representation as to the
performance characteristics of this product if it is used in conjunction with
other accessories or other than intended. If there are doubts about how to
install or clean the Laser Leveler, the effect of cleaning, the functions and/or
safety of the unit, then the unit should be withdrawn from service and the
distributor and/or manufacturer consulted (see section 11 for contact
information).