8
G-Probe
®
Device Operator Manual
13105-EN Rev E
3
Troubleshooting
General Problems
Problem
User Action(s)
No display
• Verify that the keyswitch is on.
• Verify that the components are properly connected.
• Verify that the electrical service is on.
• Inspect the fuses.
If there is still no display, contact your local IRIDEX Technical Support
representative.
Inadequate or no aiming beam
• Verify that the delivery device is properly connected.
• Verify that the console is in Treat mode.
• Turn the aiming beam control fully clockwise.
• Verify that the fiber-optic connector is not damaged.
• If possible, connect another IRIDEX delivery device and place the
console in Treat mode.
If the aiming beam is still not visible, contact your local IRIDEX Technical
Support representative.
No treatment beam
• Verify that the remote interlock has not been activated.
• Verify that the aiming beam is visible.
• Verify that the fiber switch is in the correct position for the laser system
and wavelength you are using.
• Verify that the eye safety filter is in the closed position.
If there is still no treatment beam, contact your local IRIDEX Technical
Support representative.