INW - TempHion Instruction Manual
23
Redox (ORP) Channel Calibration
Note on units: The unit “Eh” refers to readings in millivolts referenced to a hydrogen
electrode. In other words, Eh represents millivolt readings that would
have been obtained if using a hydrogen electrode. The units “mV” are
direct millivolt readings from the sensor. From the program, set the
display units to either mV or Eh, whichever you will be using on your
alternate redox meter.
— For Calibration You Will Need—
• A beaker or bucket of a sample that is representative of what will be measured
with the sensor.
DO NOT USE ZOBELL SOLUTION!
• An alternate redox meter.
• Distilled or deionized water.
• Paper
towels.
— Computing Calibration Value—
Note: When calibrating a redox channel, do not use the built-in calculator in the
Settings and Calibration Window. Instead, follow the instructions below.
• Rinse sensor with distilled or deionized water, and then pat dry with clean paper
towels.
• Place the sensor in sample. (Solution must be deep enough to cover the element
cutout.)
• Using the alternate meter, take a measurement of that sample. Use either plain
mV or Eh, whichever you normally use.
• Open the Settings and Calibration window.
• Click on the redox (ORP) channel, and then enter a zero in the offset box.
• Click
OK
.
• Take two or three single readings using the Real Time monitor and note the
redox value.
• Subtract this value from the value read by the alternate meter.
— Applying Calibration —
• Re-open the Settings and Calibration window.
• Click on the redox channel, and then enter this value in the offset box.
•
Click
OK
to save the value to the sensor!!!
— Verifying Calibration —
• Take another couple single Real Time readings. These should be close to the
reading taken with the other meter.