![INW AquiStar TempHion Instruction Manual Download Page 17](http://html.mh-extra.com/html/inw/aquistar-temphion/aquistar-temphion_instruction-manual_2085809017.webp)
16
INW - TempHion Instruction Manual
TEMPERATURE
Element Type
30K ohm thermistor
Element Material
Epoxy bead/external housing
Accuracy
± 0.2° C
Resolution
0.01°
C
Units
Celsius, Fahrenheit, Kelvin
ION SPECIFIC ELECTRODE
Chloride/Bromide
Measurement Principle
Ion electrode method
Probe Material
Ag/AgCl solid-state electrode
Range
0-10,000
ppm
Typical Accuracy
± 5.0% of measured value
Resolution
0.1
ppm
Thermal Compensation
Isopotential point characterization
Calibration
1-2 point method w/ ionic strength adjustment
Reference Solution (Chloride)
ION-Plus
B
Reference Solution (Bromide)
Potassium Nitrate - (KNO
3
)
pH/REDOX
Sensor Type/Material
pH
Glass combination electrode
Redox
Platinum
ring
Ranges
pH
0-14 pH units / -538mV to 260mV
Redox
±
1200mV
Units
pH
pH,
mV
Redox
Eh (mVH), mV
Typical Accuracy
pH
± 0.2 pH units
Redox
0.1 Eh (mVH)
Resolution
pH
0.01 pH units
Redox
0.01 Eh (mVH)
Calibration
pH
1 and 2 point w/ pH buffers (4, 7, or 10)
Redox
1
point
Reference Solution
Potassium Nitrate - (KNO
3
)