Functional Verification
5-2
3800 Vital Signs Monitor Service Manual
Revised October 2004
To place the 3800 Vital Signs Monitor back in service after functional verification has been
performed, connect the 3800 Vital Signs Monitor to a wall outlet, ensuring that cables do not
present a tripping hazard.
SpO2
•
Connect to Patient Simulator (select appropriate sensor
type).
•
Verify proper SpO2 value at 84% and 96% (±2 O2%).
•
Verify proper PR value at 30 and 240 bpm (±5 %).
NBP
•
Connect Patient Simulator and take an NBP
measurement.
•
Verify proper NBP value at 120/80 (±5 bpm).
YSI Temperature
None (self-checking).
Recorder
Verify strip chart recording of waveform data followed by a
parameter snapshot.
Monitor Function
Procedure