13
How to Test Glucose Control
1.
Wash hands with soap and water, dry
thoroughly. Allow Control, vial of Test Strips
and Meter to adjust to room temperature
(68-77ºF/20-25ºC). Write date first opened
on Control Label.
2.
Check dates on Control label and Test Strip vial
label. Do not use Control if 3 months past
written opened date
or
past date printed next to
symbol. Do not use Test Strips if 120 days
past written opened date or past date printed
next to symbol. Discard out-of-date
products and use new products if either
date has passed.
3.
Swirl or invert bottle gently to mix
Control.
DO NOT SHAKE
!
4.
Remove Test Strip from vial. Recap
vial immediately. Wipe top of vial
cap with a wet tissue, dry thoroughly.
Note:
Use Strip quickly after removal
from vial. Strips that have been
left out of vial too long will
give error message when used.
Discard old Strip and use new
Strip for testing.
5.
Insert Test Strip into Test Port.
Meter turns on.