SECTION 4—CLEANING AND DISINFECTING
Part No 1130202
29
Stratos™Portable
Cleaning the Nebulizer
WARNING
To prevent possible risk of infection due to
contamination, follow these instructions:
• Clean the nebulizer before initial operations.
• Clean the nebulizer after each aerosol treatment.
• ALWAYS use fresh cleaning solution.
• DO NOT towel dry the nebulizer components after
cleaning.
NOTE: Your physician and/or an Invacare dealer may specify
certain cleaning procedures. If so, follow their recommendations. If
not, clean according to the following instructions:
1. Remove the nebulizer and mouthpiece from the tubing.
2. Disassemble nebulizer chamber by turning
counterclockwise and separating.
3. Fill two plastic containers or bowls:
A. Washing: Hot water/detergent solution.
B. Soaking: Hot water/vinegar solution (one part vinegar
to three parts water).