Invacare® Storm®⁴ Max
Information on models with electric seat angle
adjuster
– If your wheelchair is fitted with an electric
seat angle adjuster, it is also fitted with a
microswitch which automatically reduces the
speed as soon as the seat angle exceeds around
15°. To increase speed again, readjust the seat
angle to bring it into an almost vertical position.
2.5 Safety Information With Regard to Care and
Maintenance
WARNING!
Risk of death, serious injury, or damage
Incorrect repair and/or servicing of this mobility
device performed by users/caregivers or
unqualified technicians can result in death,
serious injury, or damage.
– DO NOT attempt to carry out maintenance
work that is not described in this user manual.
Such repair and/or service MUST be performed
by a qualified technician. Contact a provider
or Invacare technician.
CAUTION!
Risk of accident and loss of warranty if
maintenance is insufficient
– For reasons of safety and in order to avoid
accidents which result from unnoticed wear, it
is important that this mobility device undergoes
an inspection once every year under normal
operating conditions (see inspection plan
contained in service instructions).
– Under difficult operating conditions such as
daily travel on steep slopes, or in the case
of use in medical care cases with frequently
changing mobility device users, it would be
expedient to carry out intermediate checks on
the brakes, accessories and running gear.
– If the mobility device is to be operated on
public roads, the vehicle driver is responsible
for ensuring that it is in an operationally reliable
condition. Inadequate or neglected care and
maintenance of the mobility device will result
in a limitation of the manufacturer's liability.
16
1547067-H