![Invacare SoftCloud Ace Service Manual Download Page 9](http://html1.mh-extra.com/html/invacare/softcloud-ace/softcloud-ace_service-manual_2082610009.webp)
60129675-B
9
Invacare
®
SoftCloud
TM
family
Installing Housing
11. Plug both FPC's and the speaker cable.
12. Install the new housing. Ensure no cable or tube
is kinked or clamped.
13. Hold the unit firmly. Turn it upside down.
14. Install the five provided screws and tighten the
new housing.
4.1.3 Replacing the PCB
R
efers to spare parts kit SP1631539
Risk of damaging the PCB by electro-
static discharge
Handle PCB's only in Electrostatic Protected
Areas
T10, T20 Torx keys
Wire clippers
1. Switch off the control unit and disconnect from
mains supply.
2. Detach the top housing, the speaker cable and
both FPC's (see 4.1.2).
3. Cut the cable tie that holds tube
C
to the bottom
housing. Disconnect the three air tubes
A
,
B
,
and
C
.
4. Unplug the motor cable
D
, the power switch
cable
E
, the compressor cable
F
, and the mains
cable
G
.
5. Cut the cable tie securing the pressure sensor
H
and remove it.
Fig. 4-4
A
C
B
Fig. 4-5
D
E
H
F
G