4.2.9 Replacing the IEC Socket
Refers to spare parts kit SP1631538
Risk of damaging the PCB by electro-
static discharge
Handle PCB's only in Electrostatic Protected
Areas
PH1 Phillips key
T20 Torx key
Wire clippers
1. Switch off the control unit and disconnect from
mains supply.
2. Detach the top housing, the speaker cable and
both FPC's (see 4.2.2).
3. Unplug the power switch cable from the PCB
(see 4.2.3).
4. Unplug the yellow earth wire from its socket (see
4.2.4).
5. Install the new socket, plug the power switch
cable to the PCB, and the earth wire in the new
socket.
6. Reinstall the top housing.