Invacare® SB® 910

9 Technical Data

9.1 Dimensions bed

All measurements are stated in cm.

All angles are stated in degrees.

All measurements and angles are stated without
tolerances, on a non-extended bed.

The

 

bed

 

can

 

be

 

extended

 

15

 

or

 

20

 

cm

 

in

 

the

 

foot

 

end.

Invacare

 

reserves

 

the

 

right

 

to

 

change

 

the

 

stated

 

measurements

 

and

 

angles

 

without

 

warning.

9.2 Weights

Max. patient weight
(provided that the weight of the
mattress and the accessories do not
exceed 50 kg)

250 kg

Max. safe working load
(p accessories)

300 kg

Bed end

5,0 kg

Lifting pole

9,0 kg

Bed total weight

141,5 kg

9.3 Dimensions mattresses

Max,

height

Mattress

Min,

height

Mattress

Min,

width

Mattress

Min,

length

Mattress

SB910 -
80 x 200
cm

18 cm

12 cm

80 cm

200 cm

SB910 -
85 x 205
cm

18 cm

12 cm

85 cm

205 cm

Min. density:

Cold foam 38 kg/m3

9.4 Environmental conditions

Storage and

transportation

Operation

Temperature

-10°C to +50°C

+5°C to +40°C

Relative humidity

20% to 75%

Atmospheric
pressure

800 hPa to 1060 hPa

Be aware that when a bed has been stored under
low temperatures, it must be adjusted to operating
conditions before use.

9.5 Electrical system

Voltage
supply

Uin 230 VAC, 50 Hz (AC = Alternating current)

Maximum
current
input

Iin max. 3 Ampere

Intermit-
tent
(periodic
motor op-
eration)

10%, max. 2 min ON / 18 min OFF

Insulation
class

CLASS I equipment

Type B
Applied
Part

Applied Part complying with the specified
requirements for protection against electrical
shock according to IEC60601-1. (An applied
parts is a part of the medical equipment
which is designed to come into physical
contact with the user or parts that are likely
to be brought into contact with the user.)

Sound
level

48 dB (A)

Degree of
protection

The control unit, hand controls, external
power supply and motors are protected
according to IPX6W.

IPX6W

- The system is protected against high

pressurized water and steam projected from
any direction.

Battery
type

Lead-acid

Battery
specifica-
tion

24 V / 1.2 Ah

The bed has no isolator (main switch). If the bed
needs to be electrically disconnected, it has to be
unplugged from the mains socket.

22

1568774-H

Summary of Contents for SB 910

Page 1: ...Invacare SB 910 en Bed User Manual This manual MUST be given to the user of the product BEFORE using this product this manual MUST be read and saved for future reference...

Page 2: ...ation duplication or modification in whole or in part is prohibited without prior written permission from Invacare Trademarks are identified by and All trademarks are owned by or licensed to Invacare...

Page 3: ...ase of the backrest 13 5 8 1 Quick release backrest CPR optional 13 5 9 Integrated mattress support extension 13 5 10 Lifting pole 14 5 10 1 Positioning the lifting pole 14 5 10 2 Adjusting the handle...

Page 4: ...ided IMPORTANT Indicates a hazardous situation that could result in damage to property if it is not avoided Tips and Recommendations Gives useful tips recommendations and information for efficient tro...

Page 5: ...are of persons with small body dimensions WARNING Risk of injury or damage Inappropriate handling of cables incorrect connections and use of unauthorized equipment can cause electrical shock and produ...

Page 6: ...ng very thick or soft mattresses low density or a combination of these increases the risk A WARNING Risk of falling The user can fall over the edge and get seriously injured if the vertical distance A...

Page 7: ...ation including technical data HMI XXXXX MD XXXXXXXXXXXXXXXX XXXXXXXXXXXXXXXX XXXXXXXXXXXXXXXX XXXXXXXXXXXXX XXXXXXXXXXXXX XXXXXXXXXXXXX XXXXXXXXXXXXX XXXXXXXXXXXXX xxx kg xxx kg YYYY MM XXXXXX XXXXX_...

Page 8: ...re listed on the product label and in the technical data section in this manual The bed is not intended for psychiatric patients The bed can be emergency operated by means of the built in accumulator...

Page 9: ...trical equipment of the bed must not be dismantled or combined with other electrical equipment After each assembly check that all fittings are properly tightened and that all parts have the correct fu...

Page 10: ...bar B of the side rail away from the locking mechanism CAUTION If the side rail is not properly locked it can fall down Pull push the top bar of the side rail to ensure that the locking system is prop...

Page 11: ...y to be set 2 Press wanted function button The LED lamp A lights up when the respective function is locked 3 Check that the corresponding function on the hand control is deactivated 4 To unlock a func...

Page 12: ...nuous connection to the mains Preventive maintenance Batteries must be charged at least every 6 months for maintenance purposes We recommend the batteries to be tested at least once a year Frequent su...

Page 13: ...e backrest until it is completely down 5 8 1 Quick release backrest CPR optional The CPR function is marked with a label on the bed CPR The quick release handle A is placed under the mattress support...

Page 14: ...e bed end fitting from the frame This can be done in two steps of respectively 15 and 20 cm 3 Release the centrally located spring and pull out the mattress support extension 4 Tighten both screws aga...

Page 15: ...ced in one of the four corner tubes 5 11 1 IV Drop rod standard 1 Place the bushing B into the desired corner tube C 2 Position the IV drop rod in the desired height 3 Tighten the finger screw A 5 11...

Page 16: ...nd disinfectants used must be effective compatible with one another and must protect the materials they are used to clean Never use corrosive fluids alkalines acid etc or abrasive cleaning agents We r...

Page 17: ...ing IMPORTANT When the bed has been washed before given to a new user it has to be thoroughly inspected Inspection must be done by a trained professional For regular maintenance see Maintenance checkl...

Page 18: ...k rest function Lock release function working properly 1 7 Height positioning Sections moving correctly motor running properly Run up down in out to outer positions 2 8 Backrest positioning Sections m...

Page 19: ...hen the hi lo motor has been replaced the mattress support is leaning cannot be set to horizontal Before resetting 1 Disconnect the bed from the mains socket and leave it without power for at least 10...

Page 20: ...a page 22 Be environmentally responsible and recycle this product through your recycling facility at its end of life Disassemble the product and its components so the different materials can be separa...

Page 21: ...s not running No relay sound is heard from the control unit Hand control is defective Replace the hand control Control unit is defective Replace the control unit Control unit is in order except for on...

Page 22: ...ric pressure 800 hPa to 1060 hPa Be aware that when a bed has been stored under low temperatures it must be adjusted to operating conditions before use 9 5 Electrical system Voltage supply Uin 230 VAC...

Page 23: ...duct uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B Thi...

Page 24: ...tic site survey should be considered If the measured field strength in the location in which this product is used exceeds the applicable RF compliance level above this product should be observed to ve...

Page 25: ...800 WLAN 802 11 a n Pulse modulation b 217 Hz 0 2 0 3 9 If necessary to achieve the immunity test level the distance between the transmitting antenna and the Medical Electrical Equipment or System may...

Page 26: ...Notes...

Page 27: ...Notes...

Page 28: ...Fax 353 1 810 7085 ireland invacare com www invacare ie United Kingdom Invacare Limited Pencoed Technology Park Pencoed Bridgend CF35 5AQ Tel 44 0 1656 776 222 Fax 44 0 1656 776 220 uk invacare com w...

Reviews: