6 OPERATION
25
Manual/Electric Portable Patient Lift
Secondary Emergency Release
In cases where the primary release is either not functioning or
unreachable, a secondary emergency release may be used.
To activate the secondary release, pull up on the EMERGENCY grip and
push down on the boom at the same time.
Performing an Emergency Stop
Press the RED button on the control box in to stop the lift arms and patient from raising or lowering.
It is recommended that the primary emergency release be
used. The secondary emergency release is only a back-up
to the primary emergency release.
Pull UP on
EMERGENCY
Grip
Push DOWN on Boom
Control Box
Press IN to Stop
Boom
Rotate CLOCKWISE to
Disengage Emergency Stop