1.
Shorten or extend the loop
D
as required.
2.
Thread loop
D
through catch
A
and plastic buckle
E
until the loop is flat.
If this adjustment is not sufficient, it might be required to
re-fit the posture belt at the fixations.
Fitting the posture belt at the fixations
CAUTION!
– Thread the belt loop through both plastic
buckles to avoid the belt from loosening.
– Do the adjustments on both sides equally,
so that the buckle clasp remains in a central
position.
– Make sure that the belts do not get caught in
the spokes of a rear wheel.
1.
Thread the loop
F
through the fixation on the chair
G
and then through BOTH plastic buckles
H
.
4.8 Oxygen cylinder holder
WARNING!
Risk of injury and/or damage to the product
– The installation of the O² holder must be
carried out by a qualified technician/ dealer.
After any adjustments, repair or service and
before use, make sure all attaching hardware is
tightened securely. Otherwise injury or damage
may occur.
WARNING!
Risk of tipping
The hanging of additional load like O² cylinder
onto your wheelchair can affect the rearward
stability of your wheelchair. This can cause the
chair to tip backwards causing injury.
– We strongly recommend the use of anti-tippers
(available as an option) when using your
wheelchair with rear additional load (O² cylinder
or similar items).
WARNING!
Risk of tipping
– Do not move the wheelchair when the backrest
is in reclined position.
– We strongly recommend the use of anti-tippers
(available as an option) when using your
wheelchair with rear additional load (O² cylinder
or similar items).
WARNING!
Risk of losing the O² cylinder
If the bands are set very loose, the O² cylinder
can falling out of its holder. This can cause the O²
cylinder to fall backwards causing injury.
– Ensure that the bands are set correctly.
IMPORTANT
– Always refer to manufacturer’s
instruction/warning for the usage of the
O² cylinder.
– The connections of the O² cylinder must
be installed and set up by a qualified
technician/dealer.
A
B
B
B
A
B
1.
Loosen the hook and loop strips
A
on the rear of the
O² holder by simply pulling on them.
2.
Insert the O² cylinder
B
into its holder.
3.
Tighten or loosen the tapes as required then reattach
them.
See section “Technical Data” for information
about recommended dimensions of the O²
cylinder.
1536533-F
21