SECTION 4—TYPICAL PRODUCT PARAMETERS
14
*Oxygen Output
Concentration Levels:
All 5LXO2AWQ/5LXAW/
5LXO2AW/9LXO2AWQ
models
* (Stated concentration levels
achieved after initial warm-up
period (approximately 30
minutes).
IRC5LXAW, IRC5LXO2AW and
IRC5LXO2AWQ
93% minimum at 1 to 3 L/min.
91% minimum at 4 L/min.
87% minimum at 5 L/min. (maximum
recommended flow)
IRC9LXO2AWQ
92% minimum at 1 to 2 L/min.
91% minimum at 8 L/min.
87% minimum at 9 L/min.
93% minimum at 3 to 7 L/min.
Maximum Outlet Pressure:
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
34.5 ± 3.45 kPa (5 ± 0.5 psi)
IRC9LXO2AWQ
62.0 kPa ± 3.45 kPa (9 ± 0.5 psi)
Flow Range:
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
0.5 to 5 L/min. Flowrates less than 1 L/min. are
not recommend
IRC9LXO2AWQ
1 L/min. to 9 L/min. Flowrates less than 1
L/min. are not recommended.
Potential Obstruction Alert:
The concentrator detects a condition that may
indicate a potential obstruction of the output
oxygen.
Rapid audible beeping alert (this alert is
deactivated when accessories are connected).
May be associated with flow setting of 0.5
L/min or less.
Average Power Consumption: (IRC5LXO2AWQ)
300 W
280 W @ 3 L/min.
(IRC5LXO2AW)
365 W
340 W @ 3 L/min.
380 W (IRC5LXAW)
460 W (IRC9LXO2AWQ)
Pressure Relief Mechanism
Operational at:
241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)
Change in maximum
recommended flow when back
pressure of 7k Pa is applied:
0.7 L/min.
Filters:
Cabinet (2), Outlet HEPA and Compressor
Inlet.
>
>