TYPICAL PRODUCT PARAMETERS
Perfecto
2
™
8
TYPICAL PRODUCT PARAMETERS
Electrical Requirements:
120 VAC ± 10% (132 VAC/108 VAC), 60 Hz
Rated Current Input:
Perfecto
2
5: 3 A
Sound Level:
Perfecto
2
5: 44 ± 2 dBA Average
Altitude:
Perfecto
2
5 - Up to 6,000 ft. (1828 meters) above sea level with-
out degradation of concentration levels.
*Oxygen Output
Concentration Levels:
All IRC5PO2/IRC5P models 95.6% to 87% at 0.5 to 5 L/min
Maximum Outlet Pressure:
Perfecto
2
5: 5 psi ± 0.5 psi. (34.5 kPa ± 3.45 kPa.)
Flow Range:
Perfecto
2
5 - 0.5 to 5 L/min
For flowrates less than 1 L/min, we recommend the use of the
Invacare Pediatric Flowmeter IRCPF16 Accessory.
Potential Obstruction Alert:
The concentrator detects a condition that may indicate a poten-
tial obstruction of the output oxygen.
Rapid audible beeping alert (this alert is deactivated when acces-
sories are connected). May be associated with flow setting of 0.5
L/min or less.
Average Power
Consumption:
Perfecto
2
5: 280 W (Operating @ 3 L/min)
Pressure Relief Mechanism Operational at:
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)
Change in maximum recommended flow
when back pressure of 7kPa is applied:
0.7 L/min
Filters:
Cabinet, Outlet HEPA and Compressor Inlet
Safety System:
Current overload or line surge shutdown. High temperature
compressor shutdown. High Pressure Alarm w/compressor shut-
down. Low Pressure Alarm w/compressor shutdown. Battery
Free Power Loss Alarm. SensO
2
Oxygen System Possible
Obstruction Alert.
Width:
15 in. ± 3/8 in (38.1 cm ± 1 cm)
Height:
23 in. ± 3/8 in (58.4 cm ± 1 cm)
Depth:
12 in. ± 3/8 in (30.5 cm ± 1 cm)
Weight:
Perfecto
2
5: 43 lbs ± 2 lbs (19.5 kg ± 1 kg)
Shipping Weight:
Perfecto
2
5: 48 lbs ± 2 lbs (21.8 kg ± 1 kg)
Operating Ambient
Temperature:
50°F - 95°F (10°C - 35°C) at 20-60% relative humidity
Exhaust:
Perfecto
2
5: Less than Ambient
+ 35°F (+19°C).
Oxygen Output:
Perfecto
2
5: Less than A 6°F (+3°C)
Cabinet:
Impact Resistant flame-retardant plastic cabinet that conforms to
UL 94-V0.