4 OPERATION
Part No 1148136
23
Invacare® MA50 Series
Max Inflate Button
1.
Press the Max Inflate button (
) to select the Max Inflate mode and inflate the mattress rapidly to maximum firmness (pressurized to 35
± 6 mmHg). A series of beeps sound every three minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the Max Inflate
mode deactivates and the control unit defaults to the previous setting.
2.
Press the Max Inflate button again to manually disengage Max Inflate mode.
Mode Button (Max Inflate/Low Air Loss)
1.
Press the Mode button (
) to select the Max Inflate mode or the Low Air Loss mode.
•
Max Inflate Mode - In this mode, the mattress inflates rapidly to maximum firmness (pressurized to 35 ± 6 mmHg). A series of beeps sound
every three minutes as a reminder that the Max Inflate mode is active. After 45 minutes, the Max Inflate mode deactivates and the control
unit defaults to the previous setting.
This button is on the MA50 and MA51 only.
It takes 20-45 minutes for the mattress to inflate fully (inflation time depends on size of mattress).
Max Inflate mode can be manually disengaged by pressing the Max Inflate button. This will deactivate the Max Inflate LED.
It is recommended that Max Inflate setting be used during patient ingress/egress, patient wound care, patient turning or patient
cleaning.
This will deactivate the Max Inflate LED.
The Mode button is on the MA55 system only.
It takes 20-45 minutes for the mattress to inflate fully (inflation time depends on size of mattress).
It is recommended that Max Inflate setting be used during patient ingress/egress, patient wound care, patient turning or patient
cleaning.