Invacare® HomeFill® II
1.
Make sure the concentrator is On. Refer to the concentrator
User Manual.
The maximum oxygen concentrator flow rate for
patient use is listed in the table below. Otherwise, the
oxygen level to the compressor will be <90% and the
compressor will not fill the cylinder.
Invacare Oxygen
Concentrator
Maximum Output Flow
for Patient Use
Platinum 5
2.5 L/min
Platinum 9
5 L/min
Perfecto
2
3 L/min
Patients with a higher prescribed flow should not refill a
cylinder while receiving supplemental oxygen treatment
from the oxygen concentrator.
2.
Set the concentrator flow rate to the patient to 2.5 L/min. or less
for Platinum 5 and 5 L/min. for Platinum 9 oxygen concentrator.
Refer to the concentrator User Manual.
3.
Make sure the cylinder is connected to the compressor. Refer
to 4.5.1 Connecting the Cylinder to the Compressor, page 17.
4.
Push the power switch
B
on the control panel
A
to the On
( ) position.
5.
Examine the control panel. The following sequence of events
should occur:
a.
0-3 minutes - The O
2
below normal (YELLOW) light
C
will
be lit while the compressor warms up.
b.
After three minutes - The filling (green) light
D
will be on
while the cylinder is filling.
c.
The full (green) light
E
will be on when the cylinder is
finished filling. Proceed to step 6.
The O
2
below normal (YELLOW) light may come on
if the oxygen level from the concentrator has not yet
reached or drops below 90% or higher. The compressor
will begin or resume filling when the oxygen level
reaches 90%. Refer to 7.2 Cylinder Fill Times, page 27 if
light does not go Off within 10 minutes.
Refer to 7.2 Cylinder Fill Times, page 27 if the Attention
(red) light
F
and audible alarm are on.
6.
Remove the full cylinder. Refer to Disconnecting the Cylinder
from the Compressor, page 18.
7.
Push the power switch to the Off (O) position.
For cylinder operation and use, refer to the specific user
manual for the cylinder/regulator assembly.
20
1145804-E
Summary of Contents for IOH200AW
Page 9: ...Safety 2 2 Label Location 1145804 E 9 ...
Page 10: ...Invacare HomeFill II A Serial Number Label B Specification Label 10 1145804 E ...
Page 35: ...Bezpečnost 2 2 Umístění štítku 1145804 E 35 ...
Page 36: ...Invacare HomeFill II A Štítek se sériovým číslem B Štítek s technickými údaji 36 1145804 E ...
Page 63: ...Sécurité 2 2 Emplacement de l étiquette 1145804 E 63 ...
Page 86: ...Notes ...
Page 93: ...Sicherheit 2 2 Anbringungsort der Etiketten 1145804 E 93 ...
Page 94: ...Invacare HomeFill II A Seriennummeretikett B Typenschild 94 1145804 E ...
Page 123: ...Sicurezza 2 2 Posizione delle etichette 1145804 E 123 ...
Page 124: ...Invacare HomeFill II A Etichetta numero di serie B Etichetta delle specifiche 124 1145804 E ...
Page 146: ...Notes ...
Page 153: ...Bezpieczeństwo 2 2 Ubicación de etiquetas 1145804 E 153 ...
Page 179: ......