SECTION 11—TIMING BELT REPLACEMENT
75
SECTION 11—TIMING BELT
REPLACEMENT
Tools Required:
7
/
16
open ended wrench,
5
/
8
open ended wrench, Diagonal Cutters, Needle Nose Pliers
Phillips screwdriver
Disassembly
NOTE: For this procedure, refer to FIGURE 11.1 through FIGURE 11.6.
1. Install an empty cylinder onto the compressor to release the pressure in the lines.
2. Remove the cylinder. Refer to Removing/Installing the Cabinet on page 25.
3. Disconnect power from the HomeFill unit.
4. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 25.
5. Disconnect the two pin connector that connects the motor to the wiring harness
6. Using needle nose pliers, disconnect the two wires from the capacitor (FIGURE 11.1).
7. Using needle nose pliers, disconnect the grounding clip from the motor
FIGURE 11.1
Two Pin
Connector
(STEP 4)
Wires (STEP 5)
Capacitor
Motor
Grounding
Clip
Control Panel
Gusset
Mounting
Screw
Control Panel
Gusset
Mounting Screw
(Hidden behind wiring assembly)