SECTION 7—BASE COMPONENTS
40
Removing Connector Fillport Dust Cover and Lanyard -
Units Manufactured Before 10/03/2005
DANGER
To prevent electrical shock, ALWAYS disconnect compressor from electrical
outlet before servicing.
Removing
NOTE: For this procedure, refer to FIGURE 7.7 on page 40.
Tools required for this procedure are:
•
7
/
16
Wrench, Phillips screwdriver
NOTE: Because the parts for the this style fillport are no longer available, remove the existing
fillport by following the instructions below and install new style fillport.
1. Remove the locknut connecting the oxygen line to the connector fillport.
2. Remove the quick-release pin.
3. Remove the mounting screw connecting the flange to the base.
4. Remove the connector fillport from the base.
5. Remove the locknut and the two washers at the bottom of the connector fillport.
6. Remove the flange from the connector fillport assembly.
7. Remove the existing lanyard from the shaft.
8. Install new style fillport. Refer to Removing/Installing Connector Fillport Dust Cover
and Lanyard - Units Manufactured After 10/02/2005 on page 41.
FIGURE 7.7
Removing Connector Fillport Dust Cover and Lanyard - Units Manufactured
Dust Cover
Shaft
Quick-Release Pin
Locknut
Washers
Locknut for
Oxygen Line
NOTE: Lanyard not shown.
Flange
Mounting Screw
Connecting Flange
to Base
Connector Fillport
Washer
Washer
Locknut