Invacare HF2RE9ES User Manual Download Page 25

Troubleshooting

7 Troubleshooting

7.1

Troubleshooting

If the Regulating Cylinder fails to function, consult the Troubleshooting guide. If problem cannot be corrected, consult your
Dealer/Home Care equipment provider.

Problem

Possible Cause

Solution

No flow.

1.

Flow dial at setting “0”.

2.

Cylinder empty.

3.

Flow dial is between settings.

4.

Kinked or blocked cannula.

5.

Shutoff valve closed.

1.

Adjust flow dial to prescribed
setting.

2.

Fill or replace cylinder.

3.

Set dial so indicator points to a
setting.

4.

Check to see if the cannula is kinked.

5.

Slowly open shutoff valve.

Cylinder will not fill.

1.

Shutoff valve closed.

1.

Slowly open shutoff valve.

1167426-E

25

Summary of Contents for HF2RE9ES

Page 1: ...ith Continuous Flow Regulator For HomeFill Oxygen System en HF2RE9ES User Manual This manual MUST be given to the user of the product BEFORE using this product read this manual and save for future ref...

Page 2: ...without prior written permission from Invacare Trademarks are identified by and All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted Making Life s...

Page 3: ...atic Testing Date 15 3 4 2 External Examination 16 4 Usage 17 4 1 Before Use 17 4 2 Connecting Disconnecting the Cylinder Assembly to from the Compressor 17 4 3 Cylinder Assembly Operation 17 5 Mainte...

Page 4: ...h if not avoided could result in death or serious injury CAUTION Caution indicates a potentially hazardous situation which if not avoided may result in property damage or minor injury or both Indicate...

Page 5: ...manual 1 3 User Profile No specific product knowledge or training is required other than what is contained in this manual to operate the product It is not intended for supplying oxygen therapy during...

Page 6: ...d conditions particular to the individual countries in which this product is sold 1 6 Component Identification Item Description A Cylinder Assembly B Cylinder C Continuous Flow Regulator D Outlet Conn...

Page 7: ...Safety 2 Safety 2 1 Label Locations Serial number located above product label shown here 1167426 E 7...

Page 8: ...d carton for damage In case of damage or if the product is not working correctly contact a technician or Invacare for repair THE INFORMATION IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE DANGER...

Page 9: ...equipment This includes sparks from static electricity created by any type of friction Keep all matches lighted cigarettes electronic cigarettes or other sources of ignition out of the room in which...

Page 10: ...ooking oil suntan oil or sunblock and similar oily products If you use these types of products Invacare suggests you wash your hands prior to using the HomeFill system If the cylinder fillport should...

Page 11: ...bly in a clean dry and cool area when not in use DANGER Risk of Injury or Death To avoid choking or ingestion of chemicals from airway contamination DO NOT use the product in the presence of pollutant...

Page 12: ...ng or kink the cannula when the product is in use as this may damage the unit DO NOT fill the cylinder assembly outdoors or anywhere below 13 C 55 F DANGER Risk of Injury or Death The cannula can caus...

Page 13: ...gnetic ferrous material it is not suitable to be in an MRI room Be sure all connections are tight and leak free DO NOT use a liquid leak detector to test for leaks Leaks are indicated by loss of press...

Page 14: ...pection section 3 3 Receiving Inspection Remove product from package and inspect for damage If there is any damage DO NOT use Contact your dealer Home Care equipment provider 3 4 Cylinder Prefill Insp...

Page 15: ...mage Cylinders that have been refinished and or exposed to elevated temperatures MUST be hydrostatically tested before filling Cylinder temperature should not exceed 60 C 140 F Aluminum cylinders subj...

Page 16: ...l or grease any other signs of damage that might cause a cylinder to be unacceptable or unsafe for use 2 Examine the cylinder for evidence of fire or thermal damage Evidence includes charring or blist...

Page 17: ...nder assembly read and follow the information in the following sections q Cylinder Prefill Inspection q External Examination 4 2 Connecting Disconnecting the Cylinder Assembly to from the Compressor R...

Page 18: ...DO NOT direct flow of oxygen at any person or flammable material when adjusting the flow dial E 4 Turn the flow dial E to setting 0 5 Remove the outlet connection cap if present 6 Attach a standard o...

Page 19: ...the flow indicator window F WARNING To avoid injury to patient ALWAYS confirm prescribed setting before administering to patient and monitor flow on a frequent basis Only use the Invacare Corporation...

Page 20: ...on section DANGER Risk of Fire Dirt or contaminants on the cylinder assembly or regulator can cause fire DO NOT use or fill cylinder assembly if the cylinder fillport is dirty or contaminated Contact...

Page 21: ...rson or flammable material when adjusting the flow dial CAUTION Risk of Damage DO NOT completely empty the cylinder Otherwise damage to the cylinder could occur 1 Position the cylinder assembly A so t...

Page 22: ...int free cloth dampened with 70 isopropyl alcohol DO NOT wipe the fillport or outlet fitting 5 Wipe the cylinder and regulator with a dry lint free cloth Ensure the product is dry before using 5 4 Dis...

Page 23: ...fillport I 4 Store product in a clean dry and cool area free from grease oil and other sources of contamination 5 7 Optional Accessories WARNING Risk of Injury or Damage Use of non Invacare accessorie...

Page 24: ...ances and recycling plans regarding disposal of the product or components normally used in operation The device does not generate waste or residue in operation DO NOT dispose of the device in the norm...

Page 25: ...ossible Cause Solution No flow 1 Flow dial at setting 0 2 Cylinder empty 3 Flow dial is between settings 4 Kinked or blocked cannula 5 Shutoff valve closed 1 Adjust flow dial to prescribed setting 2 F...

Page 26: ...inder assembly is filled with USP 93 3 oxygen using its proprietary fill port that allows filling only through the use of an Invacare HomeFill System The duration of oxygen flow between fills is depen...

Page 27: ...1 7 L 0 061 ft3 Oxygen Capacity 255 L 9 ft3 Fill Times 2 h 20 min With Concentrator flow rates to patient Concentrator Model Concentrator Flow Rate Platinum 5 Platinum S Series and Perfecto2 Series Up...

Page 28: ...Notes...

Page 29: ...Notes...

Page 30: ...Notes...

Page 31: ...San s r l Via dei Pini 62 I 36016 Thiene VI Tel 39 0445 38 00 59 Fax 39 0445 38 00 34 italia invacare com www invacare it Nederland Invacare AE Galvanistraat 14 3 NL 6716 BZ Ede Tel 31 0 318 695 757...

Page 32: ...2 4707 www invacare com EU Representative Invacare Deutschland GmbH Kleiststra e 49 D 32457 Porta Westfalica Germany Tel 49 0 5731 754 0 Fax 49 0 5731 754 52191 Manufacturer Invacare Corp 2101 E Lake...

Reviews: