Maintenance
5.2
Emptying the Cylinder Assembly
DANGER!
Risk of Fire
– NEVER smoke in an area where oxygen is being
administered.
– NEVER use near any type of flame or
flammable/explosive substances, vapors or
atmosphere.
– DO NOT direct flow of oxygen at any person,
or flammable material when adjusting the flow
dial.
CAUTION!
Risk of Damage
– DO NOT completely empty the cylinder.
Otherwise, damage to the cylinder could occur.
1.
Position the cylinder assembly
A
so the outlet
connection
D
is pointing away from the user and any
other person(s).
2.
Turn the flow dial
E
to setting “0”.
3.
Slowly open the shutoff valve
G
fully.
Turn the shutoff valve
G
counterclockwise to
open.
4.
Turn the flow dial
E
to setting “6”.
1167426-E
21