SECTION 5—MAINTENANCE
Part No. 1141494
25
Invacare Mobile Recliners
8. Turn
cylinder
1/2
‐
turn
in
the
opposite
direction
of
STEP
7.
9. Reassemble
by
reversing
STEPS
1
‐
6.
FIGURE 5.6
Adjusting Gas Spring Recline Cylinder Assembly - Model IH6077A Only
Replacing the Release Handle and Cable Assembly -
Model IH6077A Only
NOTE:
For
this
procedure,
refer
to
FIGURE 5.7
on
page 26.
1. With
one
hand,
squeeze
release
lever
up
toward
mounting
bracket
and
remove
anchor
end
of
cable
assembly
from
slot
in
release
lever
(Detail
ʺ
B
ʺ
).
CAUTION
When removing the cable from the mounting bracket, use care to pull straight out of
"V" notch. DO NOT remove by bending the cable. Damage to the cable may occur.
2. Remove
cable
assembly
from
ʺ
V
ʺ
notch
of
the
mounting
bracket
(Detail
ʺ
B
ʺ
).
3. Remove
the
mounting
screw
that
secures
the
cable
to
the
back
frame.
4. Remove
the
mounting
screw
which
secures
the
release
handle
to
the
push
handle
(Detail
ʺ
A
ʺ
).
5. Secure
new
release
handle
to
the
push
handle
with
the
existing
mounting
screw.
Securely
tighten
(Detail
ʺ
A
ʺ
).
DETAIL “B” -
TIGHTENING THE
CABLE
DETAIL “A” -
LOOSENING THE
LOCKNUT
Gas Spring
Recline
Cylinder
Locknut
Bottom Bracket
Cylinder
Cable
Clevis Pin
Clevis Pin
Gas Spring
Recline
Cylinder
Mounting
Bolt
Back Frame
Cable
Mounting
Screw
Cable
Locknut
Back
Frame
Retaining