5 Error Codes
5.1 Error Codes
The Check O
2
Plus has self diagnostic features built into the software
to detect faulty readings outside of normal operating ranges. The
codes, descriptions and recommended actions are:
•
E01:
Oxygen measurement out of range Hi (
≥
102.0% calculated
by algorithm).
Recommended Action: Verify that the Check O
2
Plus is being used
in the correct mode (Concentrator or Calibration Verification
mode). If error code repeats; perform a calibration verification. If
error code repeats again; Call Service Provider.
•
E02:
Oxygen measurement out of range Low (
≤
-2.0% calculated
by algorithm).
Recommended Action: Verify that the Check O
2
Plus is being used
in the correct mode (Concentrator or Calibration Verification
mode). If error code repeats; perform a calibration verification. If
error code repeats again; Call Service Provider.
•
E03:
Device memory corrupt or missing.
Recommended Action: Call Service Provider.
•
E04:
Signal reading not stable.
Recommended Action: Call Service Provider.
•
E05:
Pressure measurement out of Range Hi (
≥
50 PSI).
Recommended Action: Check the pressure on a known gas source
pressure. If error code repeats; Call Service Provider.
•
E06:
Outside of operating temperature Hi (
≥
40° C).
Recommended action: The Check O
2
Plus is too hot, cool the
device closer to room temperature before use.
•
E07:
Outside of operating temperature Low (
≤
15° C).
Recommended action: The Check O
2
Plus is too cold, warm the
device closer to room temperature before use.
•
E08:
Device self check found error.
Recommended Action: Remove and replace the batteries. If error
code repeats; Call Service Provider.
12