Invacare® Carroll CS Series™
Effect
Possible cause
Verification
Corrective action
Actuator not working.
Faulty actuator.
Disconnect power supply,
swap cable from unresponsive
actuator with a functional
actuator. Reconnect power
and test functions. Faulty
actuator will not work
regardless of the connection
port.
Contact Invacare Continuing
Care Service Department
1-800-668-2337.
Actuator not working.
Faulty hand control pendant.
Pendant cable is connected,
power is on, lock-out
is off and hand control
pendant does not function.
Disconnect power supply
then replace hand pendant
with functioning unit (from
another bed, if available).
Reconnect power and test.
Contact Invacare Continuing
Care Service Department
1-800-668-2337.
Bed ends to not fit properly.
Bent brackets.
Visually ensure the brackets
are parallel to the frame.
Carefully straighten minor
deformations. Call Service
Department to replace
significantly bent brackets,
1-800-668-2337.
58
1153410-Q