Aquatec® KOGIA
4.
Replace electronic cover.
Make sure that no cables get pinched.
5.
Tighten screws of electronic cover with
1.2 Nm (± 0.5 Nm).
6.
Set up backrest to seat.
7.
Check all the functions of the bath lift.
WARNING!
– The hand control should only be replaced as
a complete component.
– Do not open or attempt to repair the hand
control.
Please contact Invacare customer service by phone
or order a new hand control using the electronic
spare parts catalog at www.invacare.eu.com.
4.4 Replacing Suction Cups
Difficulty:
●○○
Tool: —
Pos.
Designation
Article
number
Quantity
A
Pair of suction cups
15071
2
Fitting
1.
Fig. 4-3
Put rear suction cup with threaded bolt
A
on sleeve
nut
B
and screw it in.
Screw as far as the suction cup fits the orientation
on the graphic.
2.
B
C
D
A
Fig. 4-4
Put front suction cup with threaded bolt
A
on sleeve
nut
B
and screw it in.
Screw as far as the suction cup fits the orientation
on the graphic.
3.
Insert toggle
C
in corresponding holding fixture
D
.
Removal
1.
Removal takes place in reverse order.
8
1650139-A