background image

1

1. Secure the frame to the floor with the suction cups or by

standing on the frame base during removal of the seat.

2. Press the UP button on the pendant to raise the

seat all the way up (DETAIL "A").

3. Grasp the handgrip on the battery pack and pull

upwards to remove it from the motor shroud as
shown in DETAIL "B".

4. Press and hold the spring loaded chair release

button to release the seat (DETAIL "C").

5. While holding the button with one hand, grasp the

seat bottom firmly with your other hand.

6. Slide the seat up until the slide block disengages from

the groove in the frame as shown in FIGURE 1.

7. Grasp the handgrip on the battery pack and push

downwards to install onto the motor shroud.

8. Press the DOWN button on the pendant to lower

the bath lift to the lowest position (DETAIL "A").

9. If necessary, remove the bath lift frame from the

bathtub. Refer to REMOVING THE FRAME FROM
THE BATHTUB in this instruction sheet.

FIGURE 1 - REMOVING THE SEAT

Frame

Groove

Seat Bottom

Slide
Block

LIFT THE SEAT

é

STEPS 5-6

Down Button

Up

Button

Pendant

DETAIL "A"

STEPS 1 AND 8

PULL UP TO REMOVE

é

PUSH DOWN TO

INSTALL

Battery Pack

ê

DETAIL "B"

STEPS 2 AND 7

Motor

Shroud

Handgrip

Spring Loaded Chair Release Button

Recess

Seat

DETAIL "C"

STEP 4

NOTE: Check all parts for shipping damage. In case of dam-
age, DO NOT use. Contact your Dealer for further instruction.

SAFETY SUMMARY

The following recommendations are made for the
safe replacement of the PENDANT/CABLE/CUR-
RENT LIMIT DEVICE:

GENERAL WARNINGS

DO NOT install or use this equipment without
first reading and understanding this instruc-
tion sheet. If you are unable to understand
the Warnings, Cautions or Instructions in this
Instruction Sheet and in the Owner's Operat-
ing and Maintenance Manual (Part No.
1081003) provided with the Bath Lift, contact
a healthcare professional, dealer or techni-
cal personnel before attempting to install this
kit, otherwise, injury or damage may occur.

This kit MUST be installed by an Invacare
Dealer or technical personnel.

INSTALLATION WARNINGS

After ANY adjustments, repair or service and
BEFORE use, make sure all attaching hard-
ware is tightened securely - otherwise injury
or damage may occur.

Use proper lifting techniques to avoid in-
jury. Keep back straight and bend knees
when lifting components of the bath lift.

REPLACING THE PENDANT/CABLE/
CURRENT LIMIT DEVICE

NOTE: Refer to the INSTALLATION WARNINGS in the
SAFETY SUMMARY of this instruction sheet.

Removing the Seat (FIGURE 1)

WARNING

To prevent pinching injury between seat
back and motor shroud when removing/
installing the seat, DO NOT grasp seat any-
where other than at the recess at the top
(DETAIL "C"), the opening at the bottom of
the seat or the lip at the top of the seat.

To prevent frame from tipping when remov-
ing seat, either suction the frame to the floor
or place one foot on the frame base when
removing/installing the seat.

CAUTION

The bath lift seat MUST be removed from the
top of the lifting  cycle,  otherwise, damage
to the unit may occur.

DO NOT grasp seat by side flaps, otherwise
damage to the bath lift or the bathtub may
occur.

Reliant Bath Lift (RBL250)

Pendant/Cable/Current Limit Device Replacement Kit

  Assembly, Installation and Operating Instructions

Kit No. 1085144

Summary of Contents for 1085144

Page 1: ...eading and understanding this instruc tion sheet If you are unable to understand the Warnings Cautions or Instructions in this Instruction Sheet and in the Owner s Operat ing and Maintenance Manual Pa...

Page 2: ...ing installation resulting in in jury and or damage WARNING ALWAYS remove the bath lift from the bath tub BEFORE performing this procedure Otherwise injury or damage may occur 1 Remove the bath lift f...

Page 3: ...Mounting Screws Nut Connection to Motor Shroud FIGURE 4 REMOVING THE MOTOR SHROUD RELEASING THE CABLES Removing Current Limit Device FIGURE 6 1 Remove the two 2 mounting screws and nuts se curing the...

Page 4: ...res should not pull out of the battery connector when gently tugged 8 Position the battery connector on the motor shroud with the flared end on the bottom as shown in FIG URE 7 9 Align the mounting ho...

Page 5: ...suctioncupsorby standingontheframebase 9 Test the operation of the bath lift by pressing the UP and DOWN buttons on the pendant to ensure proper installation of the NEW pendant 10 Perform one 1 of the...

Page 6: ...inside edge of the bathtub as shown in DETAIL G of FIGURE 9 WARNING The frame must be positioned so that the side flap of the seat on the entry exit side of the bathtub is fully sup ported by the side...

Page 7: ...TUB 3 Press the UP button on the pendant to raise the bath lift to the uppermost position 4 Grasp the handgrip on the battery pack and pull upwards to remove it from the motor shroud 5 Lowertheseatont...

Page 8: ...actory without our prior consent C O D shipments will be refused please prepay shipping charges LIMITATIONS AND EXCLUSIONS THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE...

Reviews: