Setting the Auto Cool Temperature
This sub-menu is displayed only on Refrigerating / Heating Circulators. It determines the bath
temperature at which refrigeration will be activated and permits more precise control when operating at
high temperatures as well as more rapid cool downs. For most applications, an Auto Cool set point 15°C
above room temperature is recommended. The Auto Cool range is from 1° to 150°C. The factory default
is 45°C.
Cool Command™ Refrigeration — -40°C 7 liter Refrigerating/Heating Circulators and 15 liter and larger
Refrigerating/Heating Circulators feature the Cool Command™ modulating refrigeration control system.
Cool Command allows the refrigeration system to turn on at a fluid temperature up to 150
q
C when the
temperature set point is changed to or below the Auto-Cool set point (150°C maximum). As a result, bath
fluid cools more quickly.
Conventional Refrigeration — -20°C 7 liter Refrigerating/Heating Circulators use a conventional
refrigeration system. The refrigeration system will turn on when the bath fluid temperature and set point
are below the Auto-Cool set point (70°C maximum).
To Access:
Press the
until AUTOCOOL is displayed.
To Change:
Touch the arrows until the desired auto cool temperature is displayed.
To Accept:
Press
,
,
, or allow the LCD to return to the main operational
display (approximately 10 seconds).
Resetting the Factory Default Values
To reset your Circulator to its original factory default values, proceed as follows:
1. Press
the key to place the unit in Standby.
2. Place the Power Switch / Circuit Breaker in the OFF position.
3. Return the Power Switch / Circuit Breaker to the ON position while pressing the
key until
“STANDBY” appears on the display.
4. Press
the key.
Touch and hold up/down
arrows to make large
changes
Touch and release to make
minor changes