Setting the High Limit Temperature
This sub-menu allows you to limit how high the temperature set point may be set. It also serves as a high
limit safety, alerting you if bath temperature rises above the high limit temperature setting. The High Limit
value may be set from +25° to +175°C / +80° to +350°F; the factory default is 175°C.
To avoid an unwanted shutdown during regular operation, the High Limit value should be set at least 5°
higher than the selected control temperature.
To Access:
Press the
key until HIGH LIMIT appears on the display.
To Change:
Touch the arrows until the desired high limit temperature is displayed.
To Accept:
Press
,
,
, or allow the LCD to return to the main operational
display (approximately 10 seconds).
Touch and hold up/down
arrows to make large
changes
Touch and release to make
minor changes