Package
ITS-100
Quick Start Guide
Hardware Overview
Right
Left
Error
Watt
Volt
Ohm
Unit
Display
ECP
Termination Mode
Bypass Mode
ETP
EUP
PoE IN
PoE Out
Bypass Mode
Termination Mode
Overview
This product is a device used to check the power consumption
available at the place where a camera is installed. The ITS-100
measures and displays the cable resistance from the receiver
to the location of the tester, and the power consumption of the
transmitter that can be used there.
As the tester can check the power supply and the cable status
where the camera is placed, it is also possible to use the ITS-100 for
installation and maintenance.
Features
■ Termination Mode
(B-Linx Termination Mode, ECP/EUP/ETP products)
Provides information about the cables used where the camera
will be installed.
Shows the power consumption (Watt) available for the measured
cable (Review of the PoE cameras that can be installed is possible).
Cable resistance (Ohm) from the receiver (Rx) to the tester.
Voltage supplied from the receiver (Rx) to the tester (The tester
measures overvoltage and undervoltage, not the supply voltage).
Shows an Error when an operation failure is likely to happen on a
PoE product (based on the cable voltage drop).
Shows the measurement information (periodic measure, shown
on the Display window).
Î
The tester does not work if it is not connected to a receiver (Rx).
The tester does not have a battery, and operates with the output
voltage of the receiver (Rx).
■ Bypass Mode (cable connection mode, PoE)
The tester is set in the middle of the PoE cable and measures the
cable performance while the camera is in operation.
Measures the power consumption (Watt) while the camera is in
operation.
Measures the cable voltage (Volt) while the camera is in operation.
Measures the power consumption of the camera when the tester
is installed on the camera side.
Shows real-time measurements (periodic measure, shown on the
Display window).
Operates when a PoE camera (PD device) is connected to the PoE
Output terminal.
Î
The tester might not work if it is not used with a PD device. (The
tester does not have a battery, and operates with the output
voltage from the PSE).
[Rev.2.0]
Specifications
Termination Mode
Bypass Mode (PoE)
Remarks
Operating
Voltage
(VDC)
12~60V
38~60V
Bypass Mode is limited to
the PoE operating range
(af/at)
Display
FND 2 digits, Red
2 digits, Red
7-segment display, 2
digits
LED
Red: Error
Yellow: Wattage (W)
Green: Voltage (V)
Orange: Resistance
(Ω)
Red: Error
Yellow: Wattage (W)
Green: Voltage (V)
Orange: Resistance
(Ω)
The LED shows the unit
corresponding to the FND
value displayed.
Error: If the tester is
out of the operating
range, the LEDs for the
measurement units
will flash in sequence
repeatedly.
Display
range
PoE Budget: 1-99W
Voltage: 38-58V
Resistance: 1-200 Ω
Error: 2 digits
PoE Budget: 1-60W
Voltage: 38-58V
Resistance: 1-200 Ω
Error: 2 digits
Input
terminal
ECP: BNC
EUP: RJ45 (red)
ETP: 2P Terminal model
PoE IN: RJ45 (metal)
PoE model
Output
terminal
None
PoE OUT: RJ45
(metal) connection to
PoE camera
Repetition
of the
Measurement
5s interval
3~5s interval
Information displayed
sequentially
Detection
function
No Smart detection
No PD function
None
The PSE recognizes
the connected PD
camera
The bypass Mode bypasses
the PoE signal to the
output, so it only measures
power, not speed
Measurement Display
The measurement value and the unit of measurement appear
sequentially (Integer only).
While the measure is being calculated, a small dot LED will be “On”
at the bottom of the FND.
Measurements above 100 will be displayed in two times (the first
time shows the hundreds, the second time shows digits from
00~99. The unit LED stays on while the measure is displayed).
Voltage (ex.: 38V display)
Resistance (ex.: 123 Ohm display)
"38" + "Volt LED"
"01" + "23" + "Ohm LED"
Error
Watt
Volt
Ohm
Error
Watt
Volt
Ohm
Error
Watt
Volt
Ohm
Error
Watt
Volt
Ohm
Error
Watt
Volt
Ohm
Fault Indication
There will be an error message if the measure is out of the
product specification range.
Error display: Display (FND) + Error LED + Unit LED
Failure
Display
LED
Over Load (99W and over)
Ou
+
Low voltage (38V and under)
Lo
+
Over voltage (58V and over)
Ou
+
Over current (1.2A and over)
Oc
+
Quick Start Guide
ITS-100
Cable & Power Budget Tester