2.5 Connections
1
Probe
Dedicated probe connection
2
LAN
LAN (Not used)
3
USB B
For printer, mouse, keyboard, Memory stick
4
USB A
For PC connection
5
HDMI
For external monitor or projector
6
In 24 V
Use only specified power supply unit UES65-240250SPA3 type
7
Trigger in / out
Cochlear implant trigger option
8
Pat. Resp.
Patient response button
9
Right
Audiometry Right output
10
Left
Audiometry Left output
11
Contra
Contra transducer output
2.6 Calibration cavities
You can use the 0.2 ml, 0.5 ml, 2.0ml and 5 ml for validity check of the probe calibration.
To perform a calibration check, select a protocol the measures a tympanogram.
Do not use an ear tip! Place the probe tip completely into the cavity. Perform the measurement. Check the
volume that was measured.
The allowed tolerance in the volume measurement is ±0.1 ml for cavities up to 2 ml and ±5% for larger
cavities. These tolerances are applicable for all probe tone frequencies.
We strongly recommend calibrating probe and contra phone at least once a year.