14
2.4.14 CALIBRATION METHODS
DEALER CALIBRATION
a. Turn indicator off
b. Press and hold
TARE
, then press
ON/OFF
c. Indicator displays
F1
d. Press
MODE
to enter dealer calibration
e. Press and hold
MODE
for
YES
and indicator will self calibrate zero
point before proceeding to the first point calibration
f. Load the mass according to the display, normally the first point
is 1/3 of the full capacity
g. Press
MODE
when display is flashing
h. Indicator displays
CAL._2
i. Press
MODE
for
YES
or
ZERO
to exit
j. Load the mass according to the display, normally the second point
is the 2/3 capacity
k. Press
MODE
when display is flashing to complete the calibration
procedure
AUTO CALIBRATION
a. Turn indicator off
b. Press and hold
MODE
, then press
ON/OFF
c. Indicator displays
CAL._1
d. Press
MODE
for
YES
and indicator will self calibrate zero point
before proceeding to the first point calibration
e. Load the mass according to the display, normally the first point
is 1/3 of the full capacity
f. Press
MODE
when display is flashing
g. Indicator displays
CAL._2
h. Press
MODE
for
YES
or
ZERO
to exit
i. Load the mass according to the display, normally the second point
is the 2/3 capacity
j. Press
MODE
when display is flashing to complete the calibration
procedure
Note
:
Auto calibration can only be performed when the span value
Is /-10% tolerance of dealer calibration.
Summary of Contents for UFM-B
Page 21: ...20 4 ELECTRICAL CIRCUITRY 4 1 SCHEMATICS...
Page 22: ...21...
Page 23: ...22...
Page 24: ...23 4 2 PCB LAYOUT FM 12 1 TOP LAYER FM 12 1 BOTTOM LAYER...
Page 25: ...24 FM 12 1 TOP OVERLAY FM 12 1 BOTTOM OVERLAY...
Page 32: ...31 5 APPENDIX...
Page 33: ...32...
Page 34: ...33...
Page 35: ...34...
Page 36: ...35...
Page 37: ...36...
Page 38: ...37...
Page 39: ...38...