9
-
Power supply of 5 V DC;
-
Minimum signal voltage per verification scale interval is 1.5
µ
V;
-
16 bits serial digital output;
-
Excitation power supply for the load cell is 5 V DC;
-
Minimum input impedance of the load cell is 85
Ω
;
-
Maximum cable length for the connection between the indicator and
the junction box or load cells (when more then one load cell is
connected) is 1 m/mm
2
.
-
The analog data processing unit is built in a closed metal box.
2.4 INTERNAL SETTINGS AND CALIBRATION METHODS
12 internal settings are available as below table.
Function
Symbol
Description
1
F1
Offset value reading
2
F2
Full LCD display segment checking
3
F3
Span value reading/Configuration Setup
4
F4
Auto power off setting
5
F5
RS-232 transmission setting
6
F6
Average function setting
7
F7
Data output setting
8
F8
Printout format setting
9
F9
Auto tare function setting
10
F19
1/30000 verification mode
11
F20
C.G.- gravity of calibration location
12
F21
U.G.-
gravity
of
location
of
installation
2.4.1 How to Enter the Functions Required
a. Press and hold
TARE
and turn scale on by pressing
ON/OFF
b. The scale shows
F1
c. Press
TARE
until the desired function number appears
d. Press
MODE
to enter selection
e. Press
TARE
to save and continue selection
f. Press
ZERO
to quit to restart the scale
2.4.2 Offset Value Reading (F1)
a. Remove all loads from platform
b. Enter
F1
through procedures as described in
2.4.1
c.
The display shows the offset value
Summary of Contents for UFM-B
Page 21: ...20 4 ELECTRICAL CIRCUITRY 4 1 SCHEMATICS...
Page 22: ...21...
Page 23: ...22...
Page 24: ...23 4 2 PCB LAYOUT FM 12 1 TOP LAYER FM 12 1 BOTTOM LAYER...
Page 25: ...24 FM 12 1 TOP OVERLAY FM 12 1 BOTTOM OVERLAY...
Page 32: ...31 5 APPENDIX...
Page 33: ...32...
Page 34: ...33...
Page 35: ...34...
Page 36: ...35...
Page 37: ...36...
Page 38: ...37...
Page 39: ...38...