balance with a co
ntainer on the platter to enable “Net”
weighing.
6. Backlight
a.
Press the “BL” backlight button to turn the backlight on and off.
7. Zero-Tracking and Auto-Tare
a.
Turn the balance off. Press and hold the “CAL” key and at the
same time, turn the balance on. The
display will show “-Zero-”.
Press the “TARE” key. The display will show “Zero*d”. Press
the “TARE” key to change the value from 0 to 5, “Zero0d”
meaning no zero-tracking and the sensitivity is the highest.
“Zero5d” is the factory setting. Press backlight to return to
weighing.
b.
Press the “CAL” key. The display will show “-tArE-“. Press the
“TARE” key to adjust the “*” from 0 to 9. “tArE0d” means there
is no auto-
tare. “tArE9d” is the factory setting.
c.
“Press the “BL” backlight key to confirm settings. The display
will then show “0.0”. Your settings will remain the same when
the balance is turned on again.
8. Other Functions
a.
Units.
Press “MODE” for 3 seconds. “Unit” will flash. Then
press the “TARE” key and the current unit will flash. Press the
“MODE” key to toggle between units. Press the “TARE” key to
confirm. The balance will return to weighing in the unit chosen.
b.
Counting.
Press the “MODE” key for 3 seconds until the
display flashes.
Press the “MODE” key again to choose
“Count”. Then press the “TARE” key to confirm. The display
will show “pcs”. Press the “MODE” key again to choose from
“10pcs” to “500pcs”. The larger the number, the higher the
count accuracy. Put the sample parts on the balance pan and
press the “TARE” key to confirm the sample size. The display
will show “----“. If the sample size is too small the display will
show “no-Cou”.
c.
Percentage.
Press the “MODE” key for 3 seconds until the
display flashes. Press the “MODE” key again to choose “PER”.
Then press the “TARE” key to confirm. The display will show
“100%”. Place the sample to be weighed on the pan and press
the “TARE” key to confirm the sample as 100%. Remove the
sample and place the goods on the pan. The display will show
“###%”. If the amount on the pan is too small a weight the
display will show “no-PER”.