Fiber Optic Headlight Cables
Follow all applicable bloodborne pathogen procedures as indicated by OSHA and /or your hospital requirements when cleaning,
disinfecting, and sterilizing instruments and accessories.
Cleaning (UltraLite® headlight cables*):
Wipe the fiber optic cable sheathing with a water dampened paper towel, cloth, or gauze before using 70 % Alcohol as
the disinfectant.
Disinfecting (UltraLite headlight cables*):
Soaking the UltraLite and Sweatband headlight cables in any sterilizing disinfectant should be avoided. There is a lens assembly
at the light source end, which is not sealed and could allow moisture to penetrate behind the lens.
Sterilization (UltraLite headlight cables*):
The UltraLite fiber optic cable has not been specifically designed to withstand sterilization. There is a lens assembly at
the light source end, which is not sealed, and could allow moisture to penetrate behind the lensing.
Meticulous care should be taken to avoid cable contact with sharp or pointed objects.
Maintenance
* UltraLite and Sweatband headlight cables can be identified by the curved lens at the light source end.