3
CAUTION: FEDERAL (USA) LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN
Indications for use
This fiber optic headlight cable is designed to deliver illumination from a high intensity light source to a surgical headlight
for surgical site illumination, either standard or minimally invasive. It is compatible for use with light sources made with
the following: Xenon with lamp power rating of up to 300 Watts, Halogen with lamp power rating of up to 250 Watts, or
Metal Halide with lamp power rating of up to 100 Watts. Any light source used with this cable should have a minimum of
90% IR filtering to prevent cable damage during use. This device is intended to be used by professional users.
Fiber optic headlight cables should never be used in ocular surgery or in a surgical procedure requiring direct
illumination of the eye.
See instructions for use for light source being used to ensure compatibility.
This device is CE marked in accordance with the Medical Devices Directive 93/42/EEC.