EN – ENGLISH
5
Table 1 Instrument Cycle
• Remove from washer and dry completely if needed.
• Inspect all components to ensure there is no visible organic debris or residue from cleaning
agent. Repeat process if any visible debris or residue is detected.
Disinfection
Further disinfection of device components may be achieved using one of the following methods.
Method 1:
Chemical
• With device cleaned and still disassembled, wipe all surfaces with sterile gauze moistened with
70% Isopropyl Alcohol (IPA). Be sure to achieve even coverage on all surfaces.
• Assure the device stays wet with the 70% IPA for a minimum of 10 minutes.
• Dry the device with a sterile lint free cloth or blow dry with filtered pressurized air.
Method 2:
Thermal Rinse
• A Thermal Disinfection phase may be added after the rinse cycle as indicated in table 1.
The life expectancy of the MAYFIELD products is expected to be as long as 7 years.
Phase
Time in Minutes
Water Temperature
Detergent Type &
Concentration
Pre-Wash 1
04:00
Cold tap H
2
O
N/A
Enzyme Wash
04:00
Hot tap H
2
O*
Endozime® AW
Triple plus w/
A.P.A., 1:128 ratio
Wash 1
10:00
60° C
Renu-Klenz™,
1:256 ratio
Rinse 1
00:30
Hot tap H
2
O*
N/A
Thermal Rinse**
02:00
82.2° C
N/A
* Hot Tap water not to exceed 82.2° C
** Optional phase for disinfection of components
Table 1 Instrument Cycle
Summary of Contents for A2111
Page 10: ...Adaptateur pivotant Tri Star MAYFIELD Infinity XR2 A2111 Mode d emploi ...
Page 19: ...MAYFIELD Infinity Tri Star XR2 Adattatore a perno girevole A2111 Manuale di istruzioni ...
Page 28: ...MAYFIELD Infinity XR2 Tri Star Drehadapter A2111 Gebrauchsanleitung ...
Page 37: ...MAYFIELD Infinity XR2 Tri Star Adaptador de pivote giratorio A2111 Manual de instrucciones ...
Page 46: ...MAYFIELD Infinity XR2 Tri Star warteladapter A2111 Gebruikershandleiding ...