Mount the Base Unit to the NeuroGen Adaptor
1. Loosen the Base Unit Knobs to open the receptacles
completely.
2. Insert the Support Rods of the Base Unit into the
NeuroGen receptacles. The MAYFIELD NeuroGen
Adaptor provides multiple receptacles to allow for
lateral adjustment of the Base Unit. (See Figure 1.)
3. Tighten the Base Unit Knobs to secure the Base Unit in
place.
WARNING:
Before using the Base Unit, carefully read the
product insert for the correct usage and warnings.
Cleaning and Sterilization
After each use, disassemble all components of the system
and thoroughly clean using a damp cloth and neutral pH
detergent. Remove all traces of blood and debris and dry
completely.
Inspection:
Carefully inspect all parts after each cleaning to assure all
components are in proper working condition.
CAUTION:
Components should be returned for repair if wear
is noted.
Manual Wash
CAUTIONS
• Alkaline and highly acidic detergents and solutions
cause damage to the devices.
• Channels and crevices found on this device require
particular attention during cleaning.
• Pay special attention to the water quality used
throughout reprocessing. Hard water can damage
the surface of the equipment. Avoid using hard water.
Instead use purified water unless otherwise specified.
Limitations on reprocessing
• Repeated processing has minimal effects on these
devices. Product life is normally determined by wear
and damage due to use.
• It is important to have Integra NeuroSpecialists
perform routine inspections (twice yearly is
recommended). See contact information below.
INSTRUCTIONS
Containment/Transportation
• Follow health care facility protocol for safe
containment and transport to the decontamination
environment.
• It is recommended that devices are cleaned
immediately after use.
Cleaning – Manual Equipment: Water, Neutral pH Deter-
gent, Soft Bristle Brush, Towels
Method
1. Prepare neutral pH enzymatic detergent solution (e.g.
Endozime® AW Triple Plus with APA (Ruhof), 1:128 ratio)
according to detergent manufacturer’s instructions
using lukewarm tap water.
2. Prepare equipment for soaking by disassembling
removable parts and loosening connections.
3. Rinse equipment in warm water before placing into
bath.
4. Completely soak equipment in water/detergent
solution for 30 minutes maximum.
5. Clean thoroughly with a soft nylon bristle brush.
NOTE
– If possible, use a disposable brush.
6. Rinse in warm purified water until all visible substances
and residual detergent are removed.
NOTE
– Make sure
to give special attention to hard-to-reach areas.
7. Thoroughly dry equipment with soft clean towels
and use medically compressed air if needed, to dry
channels, crevices and lumens.
8. Inspect the equipment to make sure there is no visible
organic debris or residue from cleaning agent.
Repeat process if any soil is detected.
Drying
• Products should be dry at this point. If wetness or
excess liquid is detected, dry with a soft clean towel.
• Medically compressed air can be used if needed.
Summary of Contents for A1113
Page 2: ...2 EN ENGLISH This page is intentionally left blank ...
Page 3: ...3 EN ENGLISH MAYFIELD NeuroGen Adaptor A1113 Instruction Manual REF ...
Page 10: ...10 FR FRANÇAIS FR FRANÇAIS Adaptateur MAYFIELD NeuroGen A1113 Mode d emploi REF ...
Page 18: ...18 IT ITALIANO Adattatore NeuroGen MAYFIELD A1113 Manuale di istruzioni REF ...
Page 25: ...25 25 DE DEUTSCH MAYFIELD NeuroGen Adapter A1113 Gebrauchsanleitung REF ...
Page 33: ...33 33 ES ESPAÑOL Adaptador MAYFIELD NeuroGen A1113 Manual de instrucciones REF ...
Page 41: ...41 41 NL NEDERLANDS MAYFIELD NeuroGen adapter A1113 Gebruikershandleiding REF ...