Intec PolyGard 2 AT6 User Manual Download Page 9

INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com

Specifications subject to change without notice. | Tox_D_1119, Ex_D_1119, CO2_D_0320 | USA 200413 | Page 9 of 17

SC2-AT6 – UserManual

5.2  Exchange of Sensor Cartridge

Instead of the on-site calibration, the used SC can be replaced simply and conveniently by a calibrated one.

The communication of the sensor cartridge to the sensor board is continuously monitored during 

operation and results in an immediate error message on the gas controller in case of fault or interruption. 

When replacing the sensor unit, the communication of the local bus is interrupted when unplugging the 

SC connector which leads to an immediate triggering of the error message.

•  Disconnect the SC connector from the printed circuit board (error message will be activated).
•  Loosen the locknut.
•  Remove used SC.
•  Take calibrated SC out of the original packaging, check for gas type, measuring range and valid calibration 

date.

•  Insert the SC and re-tighten with lock nut.
•  Insert the SC plug into the socket at the circuit board. Check plug for proper connection.

The local bus communication is automatically established and tested. At the same time the gas type and the 

measuring range of the “new” SC are compared with the data stored in the circuit board. If they match and the 

communication is correct, the error message will be automatically acknowledged in the Gas Controller.

The yellow LED of the circuit board flashes with a pulse of 1 sec. as long as the SC connector is disconnected 

(communication error). After the local bus communication has been re-established and the conformity test has been 

successful, the LED goes into flashing mode with 3 sec. pulse duration until the sensor’s warm-up time is over.
Apply a defined gas concentration on the sensor element with the help of the gas generator and check the 

measurement signal at the analog output or at the relay outputs.
The entire cycle “Sensor Element > Sensor Cartridge > Local Bus > Sensor Board > Field Bus > Controller” is 

observed and completed.

6  Inspection and Service

Inspection, service and calibration of the sensor should be done by trained technicians at regular intervals. We 

therefore recommend concluding a service contract with INTEC Controls or one of their authorized partners.
According to EN 45544-4, inspection and service have to be executed at regular intervals. The maximum intervals 

have to be determined and respected by the person responsible for the gas warning system according to the legal 

requirements. INTEC Controls recommends employing the common inspection and service intervals as specified in 

the general regulations of the gas measuring technique. The recommended calibration intervals depend on the sensor 

element and can be read from the table “Calibration”. If there are different intervals, always observe the shortest one.
Inspections and services must be documented. The date for the next maintenance has to be affixed to the sensor.

6.1  Inspection

Gas sensors should be controlled regularly by a competent person according to EN 45544-4. The following has to be 

checked in particular:

•  Maintenance/ calibration interval not exceeded.
•  Visual inspection of the sensor including cable for damage etc.
•  Remove dust deposits, especially at the gas inlet.

•  The filter at the gas inlet has to be replaced if extremely dirty.

Summary of Contents for PolyGard 2 AT6

Page 1: ...12700 Stowe Drive Suite 100 Poway CA 92064 Ph 858 578 7887 888 GO INTEC inteccontrols com April 13 2020 Revision Toxic E Combustible P and Infrared I CO2 Gases PolyGard 2 SC2 AT6 Sensors User Manual...

Page 2: ...missioning 6 4 1 Installation of Sensor Cartridge 7 4 2 Registration of the Sensor Cartridge 7 5 Calibration 7 5 1 Calibration Process 8 5 2 Exchange of Sensor Cartridge 9 6 Inspection and Service 9 6...

Page 3: ...fic PG2 device 1 3 Special Mode See description on datasheet for the specific PG2 device 1 4 Sensor Element for Toxic Gases and Oxygen The sensor element is a sealed electro chemical cell with three e...

Page 4: ...ne silicone Corrosive substances such as halogenated hydrocarbons Catalytic poisons such as sulphur and phosphor compounds silicon compounds metal vapours Organic solvents For testing the sensor do no...

Page 5: ...1125 C Ammonia NH3 0 300 ppm 4 000 Ceiling E1125 D Ammonia NH3 0 1000 ppm 4 000 Ceiling E1193 C Chlorine Cl2 0 10 ppm Near Source Floor E1183 C Hydrogen Cyanide HCN 0 100 ppm Near Source Ceiling E1189...

Page 6: ...tage for AT6 Install the Sensor Cartridge s if not already installed ex works Check Sensor Cartridge connector for correct engagement Calibration note that new sensors are pre calibrated at the factor...

Page 7: ...ram is also available consult your account executive Prior to calibration the Sensor Cartridge must be supplied with power voltage without interruption for warm up and stabilization The warm up time d...

Page 8: ...he buffer memory The current measured value is output with the new offset factor and the offset display is updated With Save the new offset factor is written in the SC memory only then the Zero calibr...

Page 9: ...local bus communication has been re established and the conformity test has been successful the LED goes into flashing mode with 3 sec pulse duration until the sensor s warm up time is over Apply a de...

Page 10: ...ads If this interval is exceeded a digital maintenance message is generated and forwarded Performing the calibration automatically deletes this message AT6 Sensor heads If this interval is exceeded th...

Page 11: ...correct at X4 Connect correctly Connector X4 not plugged in Check the plug Wire breakage Check the wiring Green LED doesn t flash PCB no address Check PCB address address correctly PCB no field bus co...

Page 12: ...e does not claim to be complete The indicated sensitivity data are only standard values referring to new sensor elements 9 Technical Data All specifications were collected under optimal test condition...

Page 13: ...0 2 1 2 4 149 15 90 24 1 59 5 6 12 E1130 C NO2 0 30 0 5 0 1 2 25 0 2 1 2 4 149 15 90 24 1 59 5 6 12 E1130 D NO2 0 500 20 2 2 25 0 2 1 2 4 149 15 90 24 1 59 5 6 12 E1129 C NO 0 100 0 15 0 1 2 45 0 5 1...

Page 14: ...A C4H8O2 0 100 1 0 2 1 15 0 5 0 3 1 4 122 5 95 36 3 04 Floor 6 P3425 A C2H5OH 0 100 1 0 2 1 15 0 5 0 3 1 4 122 5 95 36 1 59 Floor 6 P3410 A C2H4 0 100 1 0 2 1 15 0 5 0 3 1 4 122 5 95 36 0 98 Ceiling...

Page 15: ...00 0 5000 0 10 12 10 30 20 0 20 0 1000 150 E1125 DX NH3 10 0 100 0 100 0 200 0 5000 0 10 12 10 30 20 0 20 0 1000 150 E1125 EX NH3 10 0 100 0 100 0 200 0 5000 0 10 12 10 30 20 0 20 0 1000 150 E1193 XX2...

Page 16: ...change without notice Tox_D_1119 Ex_D_1119 CO2_D_0320 USA 200413 Page 16 of 17 SC2 AT6 UserManual 10 Figures Fig 1 Printed Circuit Board with Sensor Cartridge Mounting Mounting Mounting 115 mm X3 X2...

Page 17: ...tential bonding required also e g secondary potential to earth or grounding measures must be carried out in accordance with the respective project requirements It is important to ensure that no ground...

Reviews: