SC2-AT6 – UserManual
9.4 Cross Sensitivity - Sensor Cartridge (SC2 / AT6) / Sensor Element
Ordering No.
Gas type
Alcohols
Chlorine, Cl
2
Ethanol, C
2
H
6
O
Ethylene,
C
2
H
4
Carbon
monoxide, CO
Carbon
dioxide, CO
2
Sulfur
dioxide, SO
2
Hydrogen
sulphide, H
2
S
Nitrogen
dioxide NO
2
Nitrogen
monoxide, NO
SC2- / AT6-
ppm
ppm
ppm
ppm
ppm
ppm
ppm
ppm
ppm
ppm
E1125-AX
NH
3
10/0
100/0
100/0
200/0
5000/0
10/<10
10/<20
20/<2
20/0
1000/-10
E1125-BX
NH
3
10/0
100/0
100/0
200/0
5000/0
10/<12
10/<30
20/0
20/0
1000/-150
E1125-CX
NH
3
10/0
100/0
100/0
200/0
5000/0
10/<12
10/<30
20/0
20/0
1000/-150
E1125-DX
NH
3
10/0
100/0
100/0
200/0
5000/0
10/<12
10/<30
20/0
20/0
1000/-150
E1125-EX
NH
3
10/0
100/0
100/0
200/0
5000/0
10/<12
10/<30
20/0
20/0
1000/-150
E1193-XX2
Cl
2
300/0
5/0
20/20
35/0
300/0
E1183-XX2
HCN
100/0
100/2
20/38
15/25
5/-12
35/0
100/2
E1185-BX
CH
2
O
10-18%
1-3%
E1110-XX2
CO
2/0
2000/5
5000/0
50/0.5
25/0
50/-1
50/8
100/20
E1190-XX2
O
3
5/45/4
100/0
300/0
5/0
20/10
35/0
300/0
E1196-BX
SO
2
100/0
100/1
10/0
100/-125
100/0
100/1
E1197-XX2
H
2
S
100/2
100/20
5/1
35/2
100/20
E1130-XX2
NO
2
1/1
100/0
500/0
400/0
5000/0
30/-0.6
20/-25
50/0
1000/0
E1195-XX2
O
2
5 Vol%/
Illustration: Gas concentration of interference gas / reaction of sensor
1 The table doesn’t claim to be complete. Other gases, too, can have an influence on the sensitivity. The mentioned cross sensitivity data are only reference values valid for new sensors.
2 Cross sensitivities valid for all measuring ranges of the sensor.