5.4.1 Ranges and linearization
Different types of thermocouples have differing levels of output. They
therefore have differing fields of application most often depending on the
temperature range to be measured.
What two materials make up the couple may also be of a certain importance.
We will however, have to assume that this is known to the user.
Below is a short summary of the properties of the most common types of
thermocouples. It is included to help in selecting type and range for your
application.
”K” is the most common type.
”Cold Junction” = 0°C
TYPE
mV at
200°C
mV at
600°C
mV at
1200°C
MAX temp °C
K 8.137
24.902
48.828
1370
N 5.912
20.609
43.836
1300
J
10.777
33.096
----------
760
T
9.286
---------
----------
400
S 1.440
5.237
11.947
1760
R 1.468
5.582
13.224
1760
Thermocouples are non-linear. This is corrected by the internal program of the
”3100i”.
Thermoelectric voltage is measured (in mV) and then adjusted for the cold
junction temperature according to the type specified. This adjusted voltage is
then used as input variable to a linearizing table. The table (one for each
thermocouple type) is constructed so that its contribution to the measurement
error is negligible.