Chapter: Installation
18
M16-16654-EN
Caution
Although it is possible to move the AutoX from frame to frame, realignment and
leveling each time may be time consuming, especially if your testing involves an
extended range of travel of the AutoX arms. If you perform these types of test, it is
recommended that you have a dedicated test setup of load frame and AutoX.
Equipment
•
A suitable lifting device with a rated capacity of at least twice the weight of the
AutoX. The AutoX weighs 32kg (70 lb).
•
Metric hex keys (provided).
Moving the AutoX
Caution
Position the arms at the top of the instrument before moving it. Secure the
mechanical limit stops close to the arms to prevent movement of the arms during
transportation.
To avoid internal damage caused by the counter balance weights, position the arms at
the top of the instrument.
Caution
Take care to avoid damage to the extensometer arms during transportation.
The AutoX is a precision instrument. Its construction is robust but the extensometer
arms are vulnerable during any move.
Removing the AutoX from a frame
The procedure for removal and re-installation of the AutoX varies slightly depending on
the type of frame that it is installed on. The following procedure refers to:
•
Electromechanical frames, for example 5900 and 3300 Series
•
Static hydraulic frames, for example DX, LX, HDX, KX, KPX