Minimate Pro4 & Minimate Pro6 Alarm Setup

1. Press the 

Setup

 key to access the

 Main Menu

2. Press the 

Down Arrow

 key to select the

 View/Edit Setup File

 menu option. Press the 

Enter

 key to enter this menu.

3. Press the 

Down Arrow

 key to select the 

Auxiliary I/O 

option. Press the 

Right Arrow

 key to enter the menu.

4. Press the 

Right Arrow

 key to change the 

Auxiliary I/O

 selection to

 Remote Alarm

, as shown below:

    (the bold text in brackets “

[...]

” represents the available options) 

5. Determine how long you want the 

Warning Hold Time

 and 

Alarm Hold Time

 to sound. Enter a value using 

     the 

Right Arrow

 key to access their editing fields. Press the 

Up

/

Down Arrow 

keys to change their values. Jump 

     between the fields by using the 

Right

 or 

Left Arrow

 keys. Press the 

Enter

 key to 

Save

 the edits of each selection.  

      Note: The 

Warning Hold Time

 and 

Alarm Hold Time

 have a 300 second time limit (5 minutes). When set to 0, 

                  the alarms remain indefinitely and must be reset manually. The 

Remote Alarm Controller

 is only active once 

                  the 

Start Monitor

 key is pressed on the unit.

 

6. Select the 

Alarm On

 option and press the 

Right Arrow

 key to scroll through which sensors will trigger an alarm.  

    Each key press will provide a new setting option.

7. Select the 

Vibration Level

 option and press the 

Right Arrow

 key to edit the values.

     Note: The Minimate Pro6 monitoring unit using two geophones will offer two options to set the Vibration Alarm 

                 Level, one for each geophone attached.

8. Press the 

Right Arrow

 key to move the cursor within the value field, and the 

Up/Down Arrow

 key to  

    change the value.

9. Press the 

Enter 

key to 

Accept

 the new setting, or the 

Cancel

 key to exit the menu without saving the changes.

10. Select the 

Pressure Level

 option and press the 

Right Arrow

 key to edit the values.

       Note: The Minimate Pro6 monitoring unit with a geophone and a microphone attached will provide options 

                     to edit the levels for both sensors.

11. Press the 

Right Arrow

 key to move the cursor within the value field, and the 

Up/Down Arrow

 key to   

       change the value.

12. Press the 

Enter 

key to 

Accept

 the new setting, or the 

Cancel

 key to exit the menu without saving the changes.

13. Within the 

Edit Current Setup

 menu, press the 

Enter

 key to 

Save

 the new 

Auxiliary I/O

 settings, or press the  

       

Cancel

 key to 

Exit

 the menu without saving the changes.

 

Alarm Indicators

Warning – yellow LED with a 

pulsed

 sound from the internal warning device.

Alarm – red LED with a 

constant

 sound from the internal warning device.

External Devices – external alarm devices are optional.

  