background image

18 

 

5. IC3000 Programming command table  

 

 

Command 

Characters 

ACK 

COMMAND OPERATION 

TERMI- 

NATOR 

ACK 

DEFAULT 

01# 

Beep  Telephone number 1   (Emergency Line 1) 

Beep 

None 

02# 

Beep  Telephone number 2   (Emergency Line 2) 

Beep 

None 

03# 

Beep  Telephone number 3   (Nurse Line) 

Beep 

IC Nurse Line 

04# 

Beep  Telephone number 4   (Supervisory) 

Beep 

None 

10# 

Beep  Global account number, 4 digits 

Beep 

None 

20# 

Beep 

Line 1 Reporting Method 0 = Contact ID, 1= 4+2     

Pulse 

Beep 

Contact ID 

21# 

Beep 

Line 2 Reporting Method 0 = Contact ID, 1= 4+2 

Pulse 

Beep 

Contact ID 

22# 

Beep   Line 3 Reporting Method 2= Two-Way Voice 

Beep 

Two-Way Voice 

23# 

Beep 

Line 4 Reporting Method 0 = Contact ID, 1= 4+2 

Pulse  (Line  4  =  Supervisory  &  “Shadows”  Line  1 
reporting method) 

Beep 

Contact ID 

31# 

Beep  Ringer enable, 0 = Disable, 1 = Enable 

Beep 

Disabled 

33# 

Beep 

Help arrived signalling,  
0 = Off, 1 = On 

Beep 

Off 

34# 

Beep  Nurse Button Directed to Emergency Line 

Beep 

Disabled 

38# 

Beep 

Follow-On & Call Back Function, 0 = Follow-On, 

1= Call-Back 

Beep 

Follow-On 

39# 

Beep 

Dialling Mode 
0 = Tone, 1 = Pulse 

Beep 

Tone 

40# 

Beep  Set Access Code 

Beep 

1111 

41# 

Beep 

Auto answer for Remote programming:  
(0 = Off) (1 = On) 

Beep 

On 

42# 

Beep 

Guard time delay  
0 = Instant, 1 to 30 = 1 - 30 seconds 

Beep 

5 seconds 

43# 

Beep  Test call, 0 = OFF, 1 -30  = 1 to 30 days 

Beep 

7 days 

44# 

Beep 

Mobility time  
0= Disable, 1=12 Hour, 2=24 Hour  

Beep 

Disable 

46# 

Beep 

Follow-On mode & Two-Way Period, 
10-60 = 10 to 60 minutes 

Beep 

60 Minutes 

For Emergency 

47# 

Beep  Enter “Time” for Real Time Clock 

Beep 

(0001-2359) 

48# 

Beep  Enter “Date” for Real Time Clock 

Beep 

(MMDDYYYY) 

49# 

Beep 

Supervision 
0 = disable, 1 – 30 Days 

Beep 

7 days 

54# 

Beep 

 Nurse Call out period,  

01=1 min, to 60 = 60 minutes 

Beep 

60 Minutes 

For Nurse Line 

55# 

Beep   Answer incoming Calls by Remote 

Beep 

Off 

Summary of Contents for IC 3000

Page 1: ...Table of Contents IC 3000 MANUAL Installation Programming Operation 10 24 2012 Ver 1 9...

Page 2: ...3 Console Battery 5 2 4 Pendant Smoke Detector Other Devices Low Battery 5 2 5 No Pendant Present 6 2 6 Automatic Check in Report Periodic Test Call 6 2 7 Learning Remotes 1 2 6 2 8 Communication Form...

Page 3: ...he console should not be placed close to any heat source such as direct sunlight microwave oven which can reduce signal strength The console should not be located alongside other radio transmitting de...

Page 4: ...ndant supervisory 2 NURSE button Pendant Electrical Specifications Frequency 433 92 Mhz Battery CR2032 3V 230mAh Battery life Typical life expectancy of 5 yrs at an average use rate of one activation...

Page 5: ...the Power LED is ON and the Status LED is OFF NOTE This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are...

Page 6: ...thin 1 hour the console will transmit an AC Power Fault Report to the Monitoring Station Once power has been restored and stays on for at least 1 hour the console will transmit an AC Power Restore Rep...

Page 7: ...of Step 2 press the RED pendant button Zone is occupied Desired action Within 20 seconds of Step 3 Change to a different Zone Enter a different Command different Zone number Remove current device Pres...

Page 8: ...se Call Placed If CANCEL button was pressed before or after the call was established it will cancel hang up the call If RED Emergency button console or pendant was pressed at any time during the nurse...

Page 9: ...ep 2 If the access code is not entered or validated within this time frame the system will exit the programming mode Once in Local programming mode failure to enter any Commands within a 2 minute peri...

Page 10: ...t beep in the handset cell phone Next enter the telephone number example 18001234567 followed by the sign An audible beep will be heard This will indicate that the telephone number has been programmed...

Page 11: ...and 04 0 0 Contact ID 1 Pulse 4 2 If Contact ID is selected data is transmitted to the Monitoring Station via the ADEMCO Contact ID Standard If Pulse 4 2 is selected data is transmitted to the Monitor...

Page 12: ...ed Command 38 Follow On Call Back Function Command 38 is used to determine mode of communication after the alarm event has been received by the Monitoring Station When this function is programmed as 0...

Page 13: ...that the unit is in remote programming mode If the system responds with a Nack low tone this means that the system did not accept the access code Re enter the code Command 42 Guard Time Delay The IC...

Page 14: ...t 10 60 10 60 minutes Factory default is set as 60 60 minutes During the two way voice communication period the IC 3000 will emit two beeps via the telephone handset to alert the Monitoring Station th...

Page 15: ...See Section 2 1 for answering hanging up instructions Learning Removing Pendants and Devices The IC3000 can support a maximum of 10 zones composed of pendants devices Cmd Functions 60 Remove all zone...

Page 16: ...n be learned in are PIR Sensor PIR Door Window Sensor DW Smoke Detector SD Carbon Monoxide CO Fixed Panic Button PB Pendant Transmitter WTR Tele Medical Device TMD When programming any command if a si...

Page 17: ...ore u When the IC 3000 Low Battery status is restored to the required backup power level the Status LED will stop flashing Amber and the system will return to the Idle Mode l E333 No WTR Learned In u...