background image

New Patient Programme (Continuous)

  Set required rate; press enter button (4)

  Set volume; press enter button (4)

  Screen will now confirm all details entered; press enter button (4)

  Pump on Hold if all details correct; double press start/stop button (5) to start the pump

New Patient Programme (25 steps)

  Set bag volume; press enter button (4)

   Set number of periods; press enter button (4)

  Set infusion time for each period press enter button (4)

   Set total volume for each period press enter button (4)

  Delay time insert time if required or; press enter button (4)

  Screen will now confirm all details entered; press enter button (4)

  Pump on Hold if all details correct; double press start/stop button (5)  

to start the pump

Intermittent modes allows for dose cycle of infusions associated with HPN.

New Bag 

Confirmation screen will appear with the information from the previous bag,  
if this is correct; press the enter button (4).

  Delay time insert time if required or; press enter button (4)

  Pump on Hold; double press start/stop button (5) to start pump

Resume after Power Off

  Pump on Hold double; press start/stop button (5) to restart pump

Mini Rythmic

 PN+ 

5

Summary of Contents for Mini Rythmic PN+

Page 1: ...Training videos are also available please contact your hospital for further information Mini Rythmic PN Set up Guide...

Page 2: ...Hospital Nutrition Team Telephone No Homecare Company Telephone No Current Pump Serial No Inspiration Healthcare 2...

Page 3: ...e been installed Auto Ramp 25 Steps Continuous Volume Over Time Intermittent Mode Enter Code 0112 1 Purge button used to prime non gravity giving set 2 On off button one button with dual action 3 Soft...

Page 4: ...confirm all details entered press enter button 4 Pump on Hold if details are correct double press start stop button 5 to start the pump New Patient Programme Auto Ramp Set bag volume press enter butto...

Page 5: ...press enter button 4 Delay time insert time if required or press enter button 4 Screen will now confirm all details entered press enter button 4 Pump on Hold if all details correct double press start...

Page 6: ...iming 1 Spike the bag keep filter vertical any air will be eliminated via the filter 2 Press priming button do not use a sharp object until filter is saturated 3 Turn the priming button upside down co...

Page 7: ...k Mini Rythmic PN Pump Priming Close clamp prior to spiking the bag Load giving set into pump and follow pump programming until pump on hold Hold the pump in a vertical position with the blue clamp at...

Page 8: ...tery Pack The External Battery Pack should be connected to the pump every time it is used even when priming the giving set The life of a battery pack fully charged when used with a Mini Rythmic PN is...

Page 9: ...n menu to volume remaining or time remaining or turned off Low External Battery Pack Alarm will sound 40 minutes before the battery is completely depleted The Internal battery will AUTOMATICALLY take...

Page 10: ...ed Downstream Occlusion Indicates there is a problem between the pump and the patient Press enter button 4 to silence the alarm Rectify the problem and pump will automatically restart Upstream Occlusi...

Page 11: ...This indicates either the internal door or external casing is not closed properly This alarm will also appear if an attempt has been made to start the pump with no giving set in place Press enter but...

Page 12: ...in this leaflet are correct at time of going to press The company reserves the right to improve the products shown 500 402 7 2017 05 T 44 0 1455 840 555 E info inspiration healthcare com W inspiratio...

Reviews: