background image

Inspiration Healthcare

Keep Vein Open 

Alarm will sound when pump goes into Keep Vein Open mode. 

   Press enter button (4) to silence the alarm

  Keep Vein Open will only function if the bag volume is greater than the VTBI 

Pump Unattended

Intermittent alarm will sound when the pump is on hold or no button  
has been pressed or pump started. This will last for 1 hour.

  Press enter button (4) confirm settings and start infusion

Alarms

When there is an immediate problem the pump will Alarm and stop functioning until 
the problem has been resolved.

Downstream Occlusion

Indicates there is a problem between the pump and the patient.

  Press enter button (4) to silence the alarm  

Rectify the problem and pump will automatically restart

Upstream Occlusion or Air in Line

Indicates there is a problem between the bag and the pump.

  Press enter button (4) to silence the alarm; rectify the problem then restart  

by pressing stop/start button twice (5)

Depleted Internal Battery

Indicates the internal battery requires renewing.

  Press enter button (4) to silence the alarm.  

Turn the pump off using the on/off button (2)  
and replace the internal battery

  Turn the pump on using the on/off button (2) and restart pump by selecting  

resume after power off programme and pressing stop/start button twice (5)

10

Summary of Contents for Mini Rythmic PN+

Page 1: ...Training videos are also available please contact your hospital for further information Mini Rythmic PN Set up Guide...

Page 2: ...Hospital Nutrition Team Telephone No Homecare Company Telephone No Current Pump Serial No Inspiration Healthcare 2...

Page 3: ...e been installed Auto Ramp 25 Steps Continuous Volume Over Time Intermittent Mode Enter Code 0112 1 Purge button used to prime non gravity giving set 2 On off button one button with dual action 3 Soft...

Page 4: ...confirm all details entered press enter button 4 Pump on Hold if details are correct double press start stop button 5 to start the pump New Patient Programme Auto Ramp Set bag volume press enter butto...

Page 5: ...press enter button 4 Delay time insert time if required or press enter button 4 Screen will now confirm all details entered press enter button 4 Pump on Hold if all details correct double press start...

Page 6: ...iming 1 Spike the bag keep filter vertical any air will be eliminated via the filter 2 Press priming button do not use a sharp object until filter is saturated 3 Turn the priming button upside down co...

Page 7: ...k Mini Rythmic PN Pump Priming Close clamp prior to spiking the bag Load giving set into pump and follow pump programming until pump on hold Hold the pump in a vertical position with the blue clamp at...

Page 8: ...tery Pack The External Battery Pack should be connected to the pump every time it is used even when priming the giving set The life of a battery pack fully charged when used with a Mini Rythmic PN is...

Page 9: ...n menu to volume remaining or time remaining or turned off Low External Battery Pack Alarm will sound 40 minutes before the battery is completely depleted The Internal battery will AUTOMATICALLY take...

Page 10: ...ed Downstream Occlusion Indicates there is a problem between the pump and the patient Press enter button 4 to silence the alarm Rectify the problem and pump will automatically restart Upstream Occlusi...

Page 11: ...This indicates either the internal door or external casing is not closed properly This alarm will also appear if an attempt has been made to start the pump with no giving set in place Press enter but...

Page 12: ...in this leaflet are correct at time of going to press The company reserves the right to improve the products shown 500 402 7 2017 05 T 44 0 1455 840 555 E info inspiration healthcare com W inspiratio...

Reviews: