1

2

Operators  should  wear  protection  such  as  a  mask  and  gloves  in 
case pieces or components break away from the unit under test.

Attention 

Whether  the  unit  is  ON  or  OFF,  DO  NOT  exceed  the  capacity  of 
the  gauge.  NEVER  exceed  150%  of  the  rated  capacity,  or  the 
torque sensor will be damaged. At 110% of the rated capacity, the 
display will flash a warning.

Description

Measure  in  line  tension  and  compression  torques  only.  DO  NOT 
attempt to measure torques at an angle to the measuring shaft – 
damage to torque sensor and/or shaft may result.

Do not attempt to repair or alter this instrument. Warranty will be 
voided and damage to the unit may result.

 

Use and store within the stated temperature and humidity ranges, 
or damage and failure may result.
If not using for extended periods of time, recharge it for every 2 to 
3  months.  Or  remove  the  batteries  to  prevent  potential  battery 
leakage from causing product damage.

There are four models of IST-DCT series bottle cap torque tester, the 
capacity and resolution is in table on the bottom of the product.