OPERATION MANUAL

ISM-M1000

Metallurgical Microscope

www

.

insize

.

com

Summary of Contents for ISM-M1000

Page 1: ...OPERATION MANUAL ISM M1000 Metallurgical Microscope www insize com ...

Page 2: ... WITH A PROPERLY GROUNDED WALL OUTLET DO NOT EXPOSE THE INSTRUMENT TO HIGH TEMPERATURES OR HUMIDITY AVOID USING THE INSTRUMENT IN EXTREMELY DUSTY LOCATIONS OPERATING TEMPERATURE 5 C TO 35 C OPERATING HUMIDITY 20 TO 80 25 C WARNING WARNING 1 2 Contents This precision instrument has been designed to require a minimum of optical and mechanical maintenance Its excellent design assures years of high qu...

Page 3: ...h all adjustments and functions before using your new microscope Components 3 4 Getting Started 1 Eyepiece 2 Switch Handspike For Observe Photograph 3 Thumbscrew 4 Trinocular 5 Up Stop 6 Table board Longitudinal Adjustment Knob 7 Brightness Control Knob 8 Table board Transverse Adjustment Knob 9 Power Switch 10 Field Diaphragm Centering Screw 11 Focusing Lens Adjustment Lever 12 Light Bulb Vertica...

Page 4: ... Figure 3 The power switch to the illuminator and the brightness control is located on the base The electrical system is fuse protected and the fuse holder located on the socket Turn on the light with the power switch see figure 2 If the light does not appear to be ON check the brightness control to see if it s on a sufficiently low setting Then adjust the brightness control until image can be obs...

Page 5: ... making fine adjustments with the fine adjustment knob Focusing movement with the fine adjustment isn t affected by using the pre focusing lever Diopter Adjustment Proper correction for individual vision is accomplished via the diopter adjustment located at the left eyepiece see figure 4 Using the 40X magnification objective bring An image into focus with your right eye only Once the image is well...

Page 6: ... Figure 9 1 Set Screw 2 Cover Board 3 Lamp 4 Lamp House 5 Lamp Socket 1 2 3 4 5 a c b This paper will allow for focusing of the lamp filament on it If the filament image is not centered in the overpass hole reduce the aperture diaphragm adjust the light bulb transverse and vertical adjustment to obtain a uniform image Adjust the focusing lens lever until a sharp image of the lamp filament is attai...

Page 7: ... instrument is dry before using Microscope eyepieces and objectives are coated They should not be wiped while dry as dirt or dust may scratch the coating It is best to remove parts from the frame prior to cleaning Always blow loose dust away first Use cotton swabs or lens tissue moistened with a lens cleaner or a small amount of alcohol then wipe the surface clean with a good quality lens tissue S...