5
The Model 15/15L sensor will report the channel status as follows:
Status (hex)
Description
0000
Normal
0001
Sensor not responding
0002
Sensor error
0003
Sensor requires a zero calibration
The Model 15/15L sensor will report TSS as the primary measurement and the secondary
measurement is undefined. The units for TSS are mg/l.
The Model 51/52 sensor will report the channel status as follows:
Status (hex)
Description
0000
Normal
0001
Sensor not responding
0002
Sensor error
0003
Sensor error
The Model 51 sensor will report pH as the primary measurement and temperature as the secondary
measurement. The units for pH are hundredths of pH and the units for temperature are tenths of °C.
The Model 52 sensor will report ORP as the primary measurement and the secondary measurement
is undefined. The units for ORP are mV with a 2000 mV bias to make all readings positive.
Below is an example of a request to read the channel 2 status and channel 2 primary and secondary
measurement registers from an analyzer with the slave address of 1.
Field Name
Example (hex)
Slave Address
01
Function
04
Starting Address Hi
00
Starting Address Lo
03
No. of Regs. Hi
00
No. of Regs. Lo
03
CRC
--
Below is an example of a response to the previous query where channel 2 is connected to a Model
10 D.O. sensor measuring 8.3 ppm at 25.0°C.
Field Name
Example (hex)
Slave Address
01
Function
04
Byte Count
06
Data Hi (Reg 3)
00
Data Lo (Reg 3)
00
Data Hi (Reg 4)
03
Data Lo (Reg 4)
3E
Data Hi (Reg 5)
00
Data Lo (Reg 5)
FA
CRC
--
Summary of Contents for MPA-48
Page 22: ...2 high order byte...
Page 30: ......
Page 31: ......
Page 32: ......
Page 33: ......
Page 34: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 41: ......
Page 42: ......
Page 43: ......
Page 44: ...Suite 107 Slidell LA 70458 InsiteIG 80 Whisperwood Blvd...
Page 45: ...1550 W Lindburg Dr Slidell LA 70458 Insite IG...
Page 46: ......
Page 47: ......
Page 48: ......
Page 49: ......
Page 50: ......
Page 51: ......