40
Chapter VIII Temperature Measurement
8.1 Temperature Monitor Instruction
The portable monitor can use the temperature probe to measure the temperature
data.
Temperature Measurement Setting
•
If you are using a disposable temperature probe, insert the temperature cable
into the socket and then connect the probe with the cable. For a reusable
temperature probe, you can insert it directly into the socket.
•
The temperature probe is inserted into the anus of the animal and the
insertion depth is determined according to the size of the animal.
Warning
Before monitoring, check whether the probe cable is normal.
Warning
The disposable temperature probe can only be used once.
Warning
Handle the temperature probe and cable carefully. When not in use, the probe and
cable should be pulled into a loose ring. If the wire inside is too tight, it will cause
damage to the cable.
8.2 Temperature Alarm Information and Prompt Information
Physiological alarm, technical alarm and prompt information that may occur in TEM
measurement are listed in the table below.
Physiological Alarm
Notification
Reason
TEMP TOO HIGH
TEMP measured value is higher
than the upper alarm limit.
TEMP TOO LOW
TEMP measured value is below
the lower alarm limit.
Technical Alarm
Notification
Reason
System
Error
5/SysErr5
Temperature module self-check
error / communication error